8.

 

Open the vent valve on the back of the Fusion. This will create a pressure differential 
sufficient to push the LN2 from the supply tank into the freezer. The vent valve will remain 
open until the Fusion freezer is filled with LN2. 

 
9.

 

As the Fusion freezer and its contents are at room temperature 

for the first fill/charge, fill 

the internal sample storage area with 20-30 liters of LN2 through the neck of the tank using 
the second LN2 supply tank with another transfer hose having a phase separator 

(Not 

supplied)

 
10.

 

Connect A/C electrical power to MVE Fusion power receptacle at the rear.  

 
11.

 

After filling the sample storage area with 20-30 liters of LN2 through the neck, install the lid 

assuring that the lid lock thru hole located in the lid aligns with the lid lock tab located on 

the Dewar neck. 

 

 

Summary of Contents for 21135284

Page 1: ...MVE Fusion Series...

Page 2: ...y Chart Personal or an authorized MVE Distributor per the MVE Fusion Technical Manual PN 20994124 Product Identification Safety 3 Display Control Panel 7 Back Panel Electrical Physical Connections 8 D...

Page 3: ...al Conductivity Gas 25 83 mW m K Heat of Vaporization Liquid 198 38 kJ kg Density 1 atm Liquid 1 782 lbs L 807 4 g L 808 6 kg m3 Liquid Nitrogen Safety Transferring LN2 and operating the MVE Fusion sh...

Page 4: ...aseous nitrogen when evaporated The two most important safety aspects to consider when handling LN2 are adequate ventilation and eye and skin protection Although nitrogen gas is non toxic it is danger...

Page 5: ...asphyxia such as headache drowsiness dizziness excitation excess salivation vomiting or unconsciousness are observed remove to fresh air If breathing has stopped give artificial respiration CALL A PHS...

Page 6: ...ng Asphyxiating Atmosphere Indicates the potential for an oxygen depleted atmosphere due to nitrogen vapor Take care to operate device in a well ventilated area Warning Electricity Indicates a potenti...

Page 7: ...Display A Liquid Crystal Display LCD shows the value of all the current conditions including Freezer Temperature and LN2 level The display also shows any current alarm conditions that may exist RTD Pr...

Page 8: ...nel Electrical Physical Connections 1 RTD Temperature Probe Connection 2 Global Alarm Contacts These connections monitor All Alarms 3 Service Serial Port 4 Service Serial Port 5 Input Power 6 Main Pow...

Page 9: ...lternate inventory systems into the MVE Fusion NOTE VERY IMPORTANT TO ADD RACKS AND BOXES BEFORE FILLING CAUTION Venting may occur which will result in a loss of cryogen inside the pressure vessel if...

Page 10: ...2 to the Fusions inlet valve using the supplied transfer hose with the cane oriented in the vertical position so the pressure relief valve can vent downwards NOTE One temperature probe will be factory...

Page 11: ...installed in the secondary sensor tube the tube should remain plugged with the factory installed plug 5 Shut off the two isolation valves located underneath the shroud on both sides of the liquefier 6...

Page 12: ...at room temperature for the first fill charge fill the internal sample storage area with 20 30 liters of LN2 through the neck of the tank using the second LN2 supply tank with another transfer hose h...

Page 13: ...e 50 PSIG inside of the storage tank As the internal chamber and storage racks come down to temperature the relief events will decrease 15 Once all the steps above have been completed the MVE Fusion c...

Page 14: ...he factory using a two point low temperature range calibration method This calibration method provides a level of accuracy of 1 8 F 1 C when operating at altitudes between 1000ft to 1500ft 305m to 457...

Page 15: ...erature is out of the allowable range REJECT T OUT OF RANGE Cryocoolers reject temperature is out of the allowable range TEMP A OUT OF RANGE RTD temperature out of the allowable range TEMP A SHORT Iss...

Page 16: ...MVE FUSION Quick Reference Guide 16 21205647 D NOTES...