5.1 Calibration
The SDTA-01 does not need a periodic maintenance, however if the output values
provided by the SDTA-01 are erratic (after verification), the MULTITEL power
monitoring system’s may be programmed to compensate the differential in
measurement.
Tools and/or equipment required for calibration:
•
AC voltage clamp-on RMS multimeter
•
Terminal or portable computer to communicate with MULTITEL power
monitoring system (Emulation of VT-100, ANSI in some case)
5.1.1 Mirador or Site Manager
1.
Check that the analog channel scale is adjusted correctly.
2.
Disconnect the wire from the connector associated with the analog channel.
3.
Short-circuit the 2 wires of the analog channel.
4.
Enter the CALIBRATION command.
5.
Choose 1 for “Analog Channels”
6.
Choose the card number and the desired analog channel.
7.
Automatically adjust the zero by typing “z”.
8.
You can manually adjust the zero with the “+” and “-” keys.
9.
Replace the connector wires in place and measure the voltage applied with the
help of a multimeter.
10.
Redo the CALIBRATION command, this time specify the value applied (the
value read by the multimeter).
11.
Automatically adjust the gain by typing “G”.
12.
You can manually adjust the gain with the “<“ and “>“ keys.
5.1.2 MXP2 188
1.
Disconnect the corresponding connector for the selected analog channel on the
back panel.
2.
Short the 2 pins of the corresponding analog channel.