25 - English
9. Appendix
Safety Guidelines
CAUTION:
The socket-outlet should be installed near the equipment and
should be easily accessible.
CAUTION:
Use a power cable that is properly grounded. Always use the
appropriate AC cord that is certified for the individual country.
Some examples are listed below:
USA………...UL
Switzerland…….SEV
Canada……..CSA Britain…………..BASE/BS
Germany
……VDE Japan…….Electric Appliance Control Act
IMPORTANT NOTICE CONCERNING POWER CORD SELECTION
The power cord set for this unit has been enclosed and been selected according to
the country of destination and must be used to prevent electric shock. Use the
following guidelines if it is necessary to replace the original cord set, or if the cord
set is not enclosed. The female receptacle of the cord set must meet IEC-60320
requirements and may look like (Figure A1 below):
FOR THE UNITED STATES AND CANADA
In the United States and Canada the male plug is a NEMA5-15 style (Figure A2),
UL Listed, and CSA Labelled. For units that are mounted on a desk or table, type
SVT or SJT cord sets may be used. Any equipment standing on the floor, only SJT
type cord sets may be used. The cord set must be selected according to the
current rating for your unit. Please consult the table below for the selection criteria
for power cords used in the United States and Canada.
Code Type Size of Conductors in Cord Maximum Current Rating of Unit
SJT
18 AWG
16 AWG
14 AWG
10 Amps
12 Amps
12 Amps
SVT
18 AWG
17 AWG
10 Amps
12 Amps