background image

Page <3>

V1.0

28/06/22

Newark.com/multicomp-pro
Farnell.com/multicomp-pro
sg.element14.com/b/multicomp-pro

No.

Meanings

1

Display for current gas concentration

2

A1 low level alarm

3

Low level alarm

4

Activation of buzzer

5

High level alarm

6

A2 high level alarm

7

Activation of shutdown timer

8

(TWA),35ppm 8-hour time-weighted average (TWA) concentration means the permissible limit for carbon monoxide 

in working environment is 35ppm.

9

Zero setting icon

10

A short-term exposure limit of 15-minute time-weighted average (STEL), i.e., the permissible limit for short-term 

work is 200ppm of carbon monoxide in air.

11

Calibration icon

12

Value display lock (hold)

13

Device lock

14

Fault icon

15

Battery life

16

Gas concentration unit

17

Celsius

18

Fahrenheit

19

Temperature display/peak level of gas concentration display

20

Peak level of gas concentration display

No. Function Of Button

1

Power switch

  

(1)  Long press the button to power on the device (when it is powered off); long press the button to power off the 

 

device (when it is powered on) (Note: the detector cannot be powered off when it is self-test.) 

(2) When setting and adjusting parameters, press the button to exit current operation screen.

2

[TEMP]/[▲] button. 

(1)  When the device is under normal testing status, the button acts as [TEMP] button. Press the button, the sensor 

temperature in Celsius will be shown, and press the button again to show the sensor temperature in Fahrenheit 

will be shown. 

(2)  When setting and adjusting parameters, the button acts as [

] button. Press the button to increase values and 

long press the button to rapidly increase values.

3

[BL]/[

] button 

(1)  When the device is under normal detection status, the button acts as [BL], press the button to control the switch 

on and off of LCD backlight. 

(2)  When setting and adjusting parameters, the button acts as [▼] button. Press the button to decrease values and 

long press the button to show the rapidly decrease values.

4

[MAX] button

 

Press [MAX] to query the peak value of air concentration since start-up, long press MAX will reset the aforesaid peak 

value.

Summary of Contents for MP780892

Page 1: ...Page 1 V1 0 28 06 22 Newark com multicomp pro Farnell com multicomp pro sg element14 com b multicomp pro Carbon Monoxide Meter User Manual Part Number MP780892...

Page 2: ...lity and extended service life It provides audible and visual alarm gas concentration readings maximum value query and temperature display suitable for carbon monoxide detection in the air of resident...

Page 3: ...o power on the device when it is powered off long press the button to power off the device when it is powered on Note the detector cannot be powered off when it is self test 2 When setting and adjusti...

Page 4: ...ncentration of toxic gas carbon monoxide in the air around the measured position The response time is T90 60S The stable reading obtained 3 4min later is considered as the measured reading If any obvi...

Page 5: ...n the upper right corner is in LCD it means the battery is sufficient when the battery icon is and keeps flashing it means the battery is insufficient it requires new batteries otherwise the device ca...

Page 6: ...uld be done in room temperature recommended 20 2 C when the device temperature is close to the room temperature recommended after more than 1 hour to reduce the error due to temperature difference Und...

Page 7: ...the screen if the data is saved successfully see Fig 12 Until then the whole calibration procedure including zero setting and calibration is completed Suggestions for selection of calibrating gas con...

Page 8: ...substantially User can set the alarm point for carbon monoxide LCD displays current carbon monoxide concentration in the air checking of highest carbon monoxide concentration in history is available W...

Page 9: ...sulting from its negligence Multicomp Pro is the registered trademark of Premier Farnell Limited 2019 VI Maintenance Instruction 1 Verification Requirements Under normal temperature and humidity the s...