background image

Page <4>

V1.0

28/06/22

Newark.com/multicomp-pro
Farnell.com/multicomp-pro
sg.element14.com/b/multicomp-pro

5

[MUTE] button 

When the user has learned the alarming condition, press the button to stop current alarm and the buzzer will stop 

working within 3 minutes. Long press the button will reactivate the buzzer.

6

[HOLD] button

 

Press [HOLD] button to lock the current displayed readings; long press [HOLD] button to enter the settings for alarm 

set points A1 and A2.

7

[SET] button 

(1) Press SET under normal detection status to start sensor self-test. 

(2)  When setting the alarm points or enter the calibration mode for zero setting and calibration, press SET to choose 

parameters and long press [SET] button to save the current set values.

III. Operation Instruction 

1. Power On/Off

A yellow button   lies on the upper-left part of panel keypad. When the instrument is under shutdown status, it is able to turn 

on the instrument with a long press on the button and, when under startup status, the instrument can be shut down with a 

long press on the button (Note: shutdown is not allowed during self-checking of gas sensor). The instrument would  

automatically shut down in case of no operation or alarm action within about 20min.

Note:

 It is required to start up the instrument and perform startup countdown in clean air until there is the gas concentration 

unit (ppm icon) and during which, keep the instrument body stable and the sensor upward. Detection should be carried out 

after countdown.

2. Detection Method

Hold the instrument, keep sensor upward and near the position to be monitored, maintain the instrument body stable for 

monitoring at least 1min, then the instrument is capable of detecting if there exists high concentration of toxic gas (carbon 

monoxide) in the air around the measured position, The response time is T90<60S. The stable reading obtained 3-4min later 

is considered as the measured reading.
If any obvious change in the ambient temperature is observed, it is recommended to restart up the instrument in clean air 

when heat balance is reached between the instrument and the environment (>1 hour).
It is required to keep the instrument body stable and the sensor upward during startup and startup countdown so as to reduce 

error or zero jump caused by temperature variation or vibration.

3. Alarm

 

When the carbon monoxide concentration exceeds low alarm point A1 or TWA, the buzzer will sound and the light will flash at 

the frequency of 1 Hz; when the carbon monoxide concentration exceeds high alarm point A2 or STEL, the buzzer will sound 

and the light will flash at the frequency of 2Hz.
When the user has confirmed the current alarming condition, the user can press [MUTE] button to stop the audible alarm. The 

buzzer will stop working temporarily and the   icon in the upper left corner of LCD will disappear. After 3 minutes, the buzzer 

will start working and the icon will appear again; long press [MUTE] button when the buzzer stops working, the buzzer can be 

reactivated and the icon will appear immediately.

4. Sensor Self-Test

(1) The user should allow the device self-test at a regular basis (recommended once/24hr to once/month, frequent self-test is 

not recommended), After start-up, press [SET] button to allow the sensor self-test for 3 minutes. 
If the sensor is broken or wire breakage, short circuit, 

FAULT

 icon will flash in the upper right corner of LCD.

Please note that the sensor self-test cannot detect CO sensitivity loss caused by lack of gas diffusion when dust or water 

droplets cover the pin holes for gas diffusion. In addition, slight loss of CO sensitivity cannot be detected by self-diagnosis.

Summary of Contents for MP780892

Page 1: ...Page 1 V1 0 28 06 22 Newark com multicomp pro Farnell com multicomp pro sg element14 com b multicomp pro Carbon Monoxide Meter User Manual Part Number MP780892...

Page 2: ...lity and extended service life It provides audible and visual alarm gas concentration readings maximum value query and temperature display suitable for carbon monoxide detection in the air of resident...

Page 3: ...o power on the device when it is powered off long press the button to power off the device when it is powered on Note the detector cannot be powered off when it is self test 2 When setting and adjusti...

Page 4: ...ncentration of toxic gas carbon monoxide in the air around the measured position The response time is T90 60S The stable reading obtained 3 4min later is considered as the measured reading If any obvi...

Page 5: ...n the upper right corner is in LCD it means the battery is sufficient when the battery icon is and keeps flashing it means the battery is insufficient it requires new batteries otherwise the device ca...

Page 6: ...uld be done in room temperature recommended 20 2 C when the device temperature is close to the room temperature recommended after more than 1 hour to reduce the error due to temperature difference Und...

Page 7: ...the screen if the data is saved successfully see Fig 12 Until then the whole calibration procedure including zero setting and calibration is completed Suggestions for selection of calibrating gas con...

Page 8: ...substantially User can set the alarm point for carbon monoxide LCD displays current carbon monoxide concentration in the air checking of highest carbon monoxide concentration in history is available W...

Page 9: ...sulting from its negligence Multicomp Pro is the registered trademark of Premier Farnell Limited 2019 VI Maintenance Instruction 1 Verification Requirements Under normal temperature and humidity the s...

Reviews: