User Manual
41
Type
CAN_H
CAN_L
TX
RX
Actual CAN_H bus signal.
Actual CAN_L bus signal.
Transmission signal on the CAN signal line.
Received signal on the CAN signal line.
Sample
Point
Turn the
M
rotary control to set the Sample point, which
is a point within a bit’s time. The oscilloscope samples
the bit level at this point. “Sample point” is represented
by the percentage of “the time from the start of the bit’s
time to the sample point time” in the “bit’s time”. The
range is 5% to 95%.
Common
Baud
Turn the
M
rotary control to select from the Baud list on
the left.
Custom
Baud
Turn the
M
rotary control to set the Baud. The range is
10,000 to 1,000,000.
Tip
: You can select the nearest value in Common Baud,
and then adjust it in this menu.
Condition
Start
Trigger on the start frame of the data frame.
Type
Type
(Bottom
menu)
Data
Trigger on the selected frame.
Remote
Error
Overload
ID
Configure
(Bottom
menu)
Format
Select Standard or Extended.
ID
Use the
M
rotary control and
Direction key
on the front panel
to set.
Data
Configure
(Bottom
menu)
Byte
Length
Set the number of bytes with the
M
rotary control. The range is 1
to 8.
Data
Set the data with the
M
rotary
control and
Direction key
on
the front panel.
ID&Data
Configure
(Bottom
menu)
Format
Select Standard or Extended.
ID
Use the
M
rotary control and
Direction key
on the front panel
to set.
Byte
Length
Set the number of bytes with the
M
rotary control. The range is 1
to 8.
Data
Set the data with the
M
rotary
control and
Direction key
on
the front panel.
End
Trigger on the end frame of the data frame.