GENERAL SPECIFICATIONS                                                                                                                   
Measuring Method                : Dual integration method 
Measuring Function            : DC Voltage, AC Voltage, Resistance, Continuity Check, 

Diode Test with Data Hold & Auto Power Off functions.             

Display                                    : LCD, max. reading of 3200 with bar graph 
Polarity                                    : Automatic (

) negative indication 

Over Range Indication        : "OL" mark on LCD readout 
Low Battery Indication        : "B" mark is displayed when the battery voltage drops 
                                                        below operating voltage 
Sampling                                : 2 times/second 
Operating Temperature      : 0

40

80%RH (non-condensing) 

Storage Temperature          : 

20

60

70%RH (non-condensing) 

Power Supply                        : LR-44 x 2     
Power Consumption            : Approx. 5mW 
Size & Weight                        : 72(H)x114(W)x22.5(D)mm, approx. 110g 
Accessories                            : Instruction Manual ………… 1,    Battery …………2 

 

 

NAME & EXPLANATION OF EACH PART                                                                                                                                         

① 

ROTARY SWITCH    : Power On/Off, Measuring Ranges   

for DCV, ACV, Resistance, Diode Test                             

72

 

and Continuity Check.                       

② 

FUNCTION KEY      : When the position of rotary switch is   

at 

Ω

)))   

+, can change from 

Ω

./ 

))) / 

+ by turns.                                                                                     

③ 

DATA HOLD KEY    : By pressing this key, the display 

            114

 

value will be hold (DH sign will appear) 
and it will be released by pressing once   
again. (DH sign will disappear). 

④ 

DISPLAY                    : LCD display for measuring value,   

annunciator, battery sign, etc. 

⑤ 

TEST LEADS            : Red is for 

  polarity and Black is for 

.   

⑥ 

LEVER SWITCH      : One touch lever switch for rolling up the test lead wires into the case. 

⑦ 

BATTERY COMPARTMENT (Rear Side) 

 

 

MEASURING RANGES & ACCURACY

                                                                                                                   

(23

℃±

5

,<80%RH non-condensing) 

Function 

Range 

Accuracy 

Input Impedance  Max. Input Voltage 

 
 

DCV 

 
 
 

320.0mV 

±

(1.3%rdg

±

3dgt) 

>1000M

Ω

 

 
 
AC/DC 500Vrms 
 
 

3.200V 

±

(0.7%rdg

±

3dgt) 

approx. 11M

Ω

 

32.00V 
320.0V 
500V 

 

±

(1.3%rdg

±

3dgt) 

 

 
approx. 10M

Ω

 

 

 

ACV 

(50/60Hz) 

3.200V 

 
    ±(2.3%rdg±6dgt) 
     
 

approx. 11M

Ω

 

 
AC/DC 500Vrms 
 

32.00V 
320.0V 
500V 

 
approx. 10M

Ω

 

 

 

Function 

Range 

Accuracy 

Testing Current  Max. Input Protection 

 
 

OHM 

 
 
 

320.0

Ω

 

±

(2%rdg

±

5dgt) 

<0.7mA 

 
 

500Vrms 

 
 
 

3.200k

Ω

 

 

±

(2%rdg

±

3dgt) 

 

<0.13mA 

32.00k

Ω

 

<13

μ

320.0k

Ω

 

<1.3

μ

3.200M

Ω

 

±

(6%rdg

±

4dgt) 

<0.13

μ

 

32.00M

Ω

 

±

(10%rdg

±

10dgt) 

   

Diode Test 

  3.200V 

±

(10%rdg

±

3dgt) 

  approx.. 0.7mA 

         

500Vrms 

Continuity 
Check 

320.0

Ω

 

Continuity Beeper, 

<approx. 20

Ω

 

  approx.. 0.7mA 

         

500Vrms

 

 

Summary of Contents for MCD-008

Page 1: ...injury or only the material damage can be supposed caused from mis operations The CAUTIONs and WARNINGs which appear on the following pages are stared to prevent the operator other people from the dangers and their properties from the damages beforehand WARNING POCKET DIGITAL MULTIMETER MODEL MCD 008 INSTRUCTION MANUAL POSSIBLE ELECTRICAL SHOCK This instrument is for the use of low voltage circuit...

Page 2: ... can change from Ω by turns DATA HOLD KEY By pressing this key the display 114 value will be hold DH sign will appear and it will be released by pressing once again DH sign will disappear DISPLAY LCD display for measuring value annunciator battery sign etc TEST LEADS Red is for polarity and Black is for LEVER SWITCH One touch lever switch for rolling up the test lead wires into the case BATTERY CO...

Page 3: ...ed wire of test leads before connecting In case of out of order stop the operation and repair them 5 POSSIBLE DAMAGE FIRE HAZARD Connect test leads firmly the wrong connection may cause spark 3 RESISTANCE Ω 1 Set the rotary switch position to Ω and chose Ω by select key 2 According to the drawing contact test leads to the both sides of the resister to be measured It does not matter about polarity ...

Page 4: ...attery first disconnect the test leads from any energized circuit Disconnect the test leads from any energized circuit Set the function switch to OFF slide Remove the battery cover screw Slide of battery cover and change two batteries Set the battery cover and screw battery cove WARRANTY This instrument is sent out from our factory after the sufficient internal inspections but if you find any defe...

Reviews: