①
Press POWER switch
⑤
once. All display appears and then, becomes voltage
display.
②
Connect ZCT to CT INPUT and test leads to
~
V INPUT according to each color.
③
According to the circuit and apparatus to be measured, input the voltage and
current to the instrument. Refer to wiring method for voltage & current input.
④
By frequency change switch, set the power supply frequency to be measured.
⑤
Confirm the voltage value on display and once press current range switch. The
display becomes Io current. Confirm Io current value and press current range
switch once again. The display becomes 1
Φ
/ Ior showing Ior current of single
phase. Measure Ior current value by pressing current range switch according to
wiring system to be measured.
By pressing current range switch :
Io I
Φ
/Ior
△
/Ior
Y/Ior
press press press
press
In order to consider tan
δ
, once press tan
δ
ON switch
⑬
, “tan
δ
” appears on the
display. To get back before, press tan
δ
OFF switch
⑭
and “tan
δ
” disappears.
⑥
To measure the insulation resistance (M
Ω
), once press insulation resistance
switch. In case of no voltage input, the display becomes “OL”.
⑦
In order to memorize Ior current or M
Ω
values, press D-HOLD switch
⑩
and
then, press ENT switch
⑪
at each time.
(WIRING SYSTEM)
○
a 1P/1W, 1P/3W CIRCUIT : Wiring Mode 1
Φ
/Ior