3. DIMENSIONS AND PANEL FUNCTION   

     

Current transducer(Jaw)

                       

 

Jaw opening lever 

   

Power switch 

   

Range selector switch 

Data hold switch 

   

LCD display 

   

Input terminal(V) 

COM terminal 

Input terminal(

Ω

)(

     

)

 

Battery cover 

11

Wrist strap 

                                           

                 

 
 
 
 
 

4. METHOD OF MEASUREMENT 
 

4-1. Measurement of AC Current (

A)   

 
4-1-1 Measurement of Line Current 

① 

Press the power switch "ON". 

② 

Set the range selector switch to the proper   

ACA range to be measured.   
Start the measurement at top range and   
then work down to lower range. 

③ 

Open the CT and clamp it to the conductor   
to be measured at the center of CT.                                               

Close CT completely. 

④ 

Read the current value on the display. 

⑤ 

If you make measurements in a dark place   

or in a place where it is difficult to see the   
readings, use the data hold switch.                                                  Fig.1 

                                                                                                                             

Note : Clamp around only one conductor of the circuit to be measured. (See Fig.1)                                                 

When making the clamp measurements for high AC current, the noise sound 
of CT may be caused, however, this is not abnormal. 

 

4-1-2 Measurement of Leakage Current 

① 

For  leakage  current  measurement,  clamp  CT  to  the  grounding  line  in  the  same 
manner as for line current measurement. 

② 

This model cannot measure zero phase current accurately by clamping CT to the 
plural conductors together (2 wires for single phase, 3 wires for 3 phase, etc.). 

 
Note  :  The  display  may  show  some  value  even  without  clamping  CT  nearby  the 

conductor and or transformer carrying big current due to the influence of outer 
magnetic fields.    In such case, operate the instrument in the place apart from 
the source of strong magnetism as much as possible. 
 

Summary of Contents for MCL-550D

Page 1: ...anical electronic design Before you use your new instrument read this Instruction Manual completely and familiarize yourself thoroughly with all functions Keep this manual carefully to take out whenev...

Page 2: ...operations WARNING POSSIBLE ELECTRICAL SHOCK This instrument is for the use of low voltage circuit Do not make measurements in the circuit more than AC 600V Before use check and confirm the voltage o...

Page 3: ...f mode approx 10 minutes after the power switch on Operating temperature 0 to 40 80 RH non condensing Strange temperature 10 to 60 70 RH non condensing Power supply 1 5V AAA size R03 x2 Power consumpt...

Page 4: ...where it is difficult to see the readings use the data hold switch Fig 1 Note Clamp around only one conductor of the circuit to be measured See Fig 1 When making the clamp measurements for high AC cur...

Page 5: ...tion Connect the plug of red test lead into the V terminal and the plug of black test lead into the COM terminal Set the range selector switch to ACV range Contact the tips of the test leads to the ci...

Page 6: ...to rain or moisture Do not operate the tester in the presence of flammable gases or fumes CAUTION To avoid damage to the tester disconnect test leads before changing functions Do not exceed the maximu...

Page 7: ...e displayed CAUTION Be sure all voltage is turned OFF in the circuit before making diode test 5 REPLACEMENT OF BATTERIES When the battery voltage falls below a specified voltage value the B mark will...

Page 8: ...ent out from our factory after the sufficient internal inspections but if you find any defect due to the fault in our workmanship or the original parts please contact the dealer where you bought the i...

Reviews: