2. SPECIFICATIONS
Safety standard : Meets the requirements for double insulation to IEC 1010-2-032,
IEC 61010-1, IEC 61010-2-032 installation Category
Ⅲ
600V
phase to earth.
E.M.C. standard : The instrument meets EN 61326-1/-2-2 and EN 61000-4-2/-3
Withstanding voltage : AC 5500V, 1 minute (between outer case and core of CT)
Measuring method : Dual integration mode
Jaw opening capability : 55mm
φ
Display : 3 1/2 digit LCD max. reading of 1999 and annunciators
Over range indication : Blanking of all digits except MSD1
Low battery indication : "B" mark on LCD readout
Sampling : 2 times/s
Data hold indication : "D
・
H" mark on LCD readout
Auto power off : The meter is set to power off mode approx. 10 minutes after the
power switch on.
Operating temperature : 0
℃
to 40
℃
,
<
80%RH (non-condensing)
Strange temperature :
‐
10
℃
to 60
℃
,
<
70%RH (non-condensing)
Power supply : 1.5V (“ AAA” size, R03) x2
Power consumption and battery life : Approx. 5mW, 500 hours continuous.
Size : 85(W)x240(H)x34(D)mm
Weight : Approx. 350g
Accessories : Carrying case…………..1
Instruction manual……1
Batteries…………………2
Test lead…………………1set
Measuring range (23
℃±
5
℃
,
<
80% RH ) :
Range
Accuracy
Max. input
~A
(50/60Hz)
Manual range
20A
±
1.2% rdg
±
10 dgt
AC2000A
200A
〃
2000A
±
1.2% rdg
±
8 dgt
~V
(50/60Hz)
… V
Auto range
2V
±
0.7% rdg
±
5 dgt
AC/DC
750V rms
20V
±
1.2% rdg
±
5 dgt
200V
〃
750V
〃
Ω
(OHM)
Auto range
200
Ω
±
1.2% rdg
±
5 dgt
Input protection
2K
Ω
〃
250V rms
20K
Ω
〃
200K
Ω
〃
2000K
Ω
〃
20M
Ω
±
3% rdg
±
10 dgt
・
)
)
)
Continuity
check
2K
Ω
Continuity beeper
<
Approx. 300
Ω
250V rms
|
Diode test
2V
±
10% rdg
±
3 dgt
250V rms