5-4. Measurement of Diode 
 

  Set the power switch to "ON" position.   

  Connect the plug of red test lead into the “   

” terminal and the plug of black   

test lead into the “COM” terminal. 

  Set the range selector switch to “   

” range. 

  Contact the tips of the test leads to the diode under test. 

 
 

 

 
 

Fig.4 

 
  As shown in Fig.4 (a), when measuring the forward voltage across the diode, a 

normal diode will indicate 0.4 to 0.7V and, as shown in Fig.4 (b) the reverse voltage 

will indicate a value near 1,500. For a short-circuited diode, a value near 0mV will be 
displayed. 

 

                                                                                                                                   

 

!

CAUTION 
Be sure all voltage is turned OFF in the circuit before making diode test.   

 
 

6. REPLACEMENT OF BATTERIES AND FUSE 

 

   

When the battery voltage falls below a specified voltage value, the "B" mark will 

appear in the display. If this occurs, replace the two batteries with new ones as 

quickly as possible.   

Turn the power switch to "OFF", prior   

to installing batteries. 

    To install the batteries, remove the   

battery cover located on the unit back. 

    Loosen the screw on the battery cover. 
    Replace the two batteries (UM-4 or   

type AAA) with new ones, observing   

polarity. Use high-quality batteries   
which are guaranteed against leakage.   

If the instrument is to be left unused   

for long periods of time, to prevent   
damage from leakage, remove the batteries. 

 

 

 

!

WARNING 
Before changing batteries or fuse, removal all electrical input.   

 

 

Summary of Contents for M-2010

Page 1: ...Before you use your new instrument read this Instruction Manual completely and familiarize yourself thoroughly with all functions With proper use and care your tester will give you years of satisfact...

Page 2: ...nd etc Before operating this instrument familiarize yourself with all instructions outlined in this manual Always check to make sure that the function switch is set to the proper position When making...

Page 3: ...LCD readout Auto power off The meter is set to power off mode approx 10 minutes after the power switch on Operating temperature 0 to 40 80 RH non condensing Strange temperature 10 to 60 70 RH non cond...

Page 4: ...a range appropriate to the current to be measured Start the measurement at top range and then work down to lower range Clamp the conductor of the circuit at the center of CT Read the current value in...

Page 5: ...onnect the plug of red test lead into the V terminal and the plug of black test lead into the COM terminal Set the range selector switch to ACV range Contact the tips of the test leads to the circuit...

Page 6: ...rain or moisture Do not operate the tester in the presence of flammable gases or fumes CAUTION To avoid damage to the tester disconnect test leads before changing functions Do not exceed the maximum i...

Page 7: ...d OFF in the circuit before making diode test 6 REPLACEMENT OF BATTERIES AND FUSE When the battery voltage falls below a specified voltage value the B mark will appear in the display If this occurs re...

Page 8: ...this instrument always pack it in its original or equivalent packing material and pack together with name address telephone number and the warranty documentation To ensure speedy and reliable repair a...

Reviews: