background image

 

 

1.  Introduction 

1.1 

Intended Use  

The  OX-700  is  intended  to  be  used  for  measuring, 
displaying  and  storing  of  oxygen  saturation  (SpO

2

 

pulse  rate  (PR)  and  motion  in  home  or  a  healthcare 
facilities environment. The App can show real-time SpO

2

pulse  rate  and  waveform,  display  results  and  share 
reports. 
 
The data and results provided by this device are for pre-
check  screening  purpose  only  and  cannot  be  directly 
used for diagnostic or medical treatment. 

 Warnings and Cautionary Advices 

·

 

DO NOT squeeze the sensor or apply excessive force on the sensor & 
cable.  

 
 
 
 

Ÿ

 

Do not use this device during MRI examination. 

Ÿ

 

Never submerge the device in water or other liquids. Do not clean the 
device with acetone or other volatile solutions.  

Ÿ

 

Do not place this device in pressure vessels or gas sterilization device. 

Ÿ

 

Consult your doctor immediately if you experience symptoms that could 
indicate acute disease.  

Ÿ

 

Do not self-diagnose or self-medicate on the basis of this device without 
consulting  your  doctor.  In  particular,  do  not  start  taking  any  new 
medication or change the type and/or dosage of any existing medication 
without prior approval.  

Ÿ

 

Use only cables, sensors and other accessories specified in this manual.  

Ÿ

 

Prolonged  continuous  SpO

2

  monitoring  may  increase  the  risk  of 

undesirable changes in skin characteristics, such as irritation, reddening, 
blistering or burns.  

Ÿ

 

Check the SpO

2

 sensor application site every 6-8 hours to determine the 

positioning of the sensor and the circulation and skin sensitivity of the 
patient.  Patient  sensitivity  varies  depending  on  medical  status  or  skin 
condition. For patients with poor peripheral blood circulation or sensitive 

Summary of Contents for OX-700

Page 1: ...Spo2 Sleep Fitness Monitor Wrist Pulse Oximeter OX 700 User s Manual...

Page 2: ...User s Manual Contents 1 Introduction 1 2 Using the Monitor 3 3 Maintenance 9 4 Troubleshooting 9 5 Specifications 10 6 Electromagnetic Compatibility 11...

Page 3: ...ot place this device in pressure vessels or gas sterilization device Consult your doctor immediately if you experience symptoms that could indicate acute disease Do not self diagnose or self medicate...

Page 4: ...d pressure cuffs infusion lines etc moisture in the sensor improperly applied sensor incorrect sensor type poor pulse quality venous pulsations anemia or low hemoglobin concentrations cardiogreen and...

Page 5: ...ith a tool per IEC 60529 Follow Instructions for Use Warning and Caution SN Serial number No alarm system 1 3 OX 700 Contains 1 x Monitor 1 x Ring Sensor 1 x Charging Cable 1 x User s Manual 2 Using t...

Page 6: ...tion and no App connection 2 4 Bluetooth Connection The device Bluetooth will be enabled automatically only when the display screen is active On To establish a Bluetooth connection 1 Ensure the device...

Page 7: ...SYNC Start the device by holding the start stop button for 2 seconds Current time will display Now start App and it will connect with the device and upload any new files Spo2 Sleep Mode Fitness Monit...

Page 8: ...tart counting after the ring sensor plugged in Please avoid excessive motion for the sensed finger during recording and avoid any strong ambient light condition 2 8 Stop monitoring sync data Take off...

Page 9: ...e display information 2 10 Real time Dashboard You can check real time SpO2 PR and pulse waveform using the App 1 Connect the device to App 2 On the App select Dashboard from the Lower Tab line 2 11 S...

Page 10: ...alarm in of Spo2 2 12 View Results and Report In the App you can review all the historic data By touching a record the App will display a detailed report It includes analysis results and charts Slide...

Page 11: ...months when the device is not in use 4 Troubleshooting Problem Possible Cause Possible Solution Device does not turn on or has no response Battery may be low Charge battery and try again Unexpected so...

Page 12: ...DC 5V 10 Battery type Rechargeable lithium polymer battery Charge time 2 hours SpO2 Standards Meet standards of ISO 80601 2 61 Measurement accuracy verification The SpO2 accuracy has been verified in...

Page 13: ...gnetic emission or decreased electromagnetic immunity of the equipment The device or its components should not be used adjacent to or stacked with other equipment The device needs special precautions...

Page 14: ...tured for Medical Test Supply LLC Tel 1 855 270 5357 E Mail admin medicaltestsupply com Website www medicaltestsupply com PN OX 700M Version B Contents of this manual are subject to changes without pr...

Reviews: