M-Series Model MG Digital Gauge, Transmitter Electronics - Replacement Guide 551104 A
Installing current parameters
5
Setting new gradient value and board parameters
Your new boards must be programmed before use. Follow the steps
below to program your new boards:
Launch your setup software and connect to the transmitter. Click
1.
the “
Data from Device
” tab, then click the “
Backup/Restore
” button.
The “
Backup and Restore Device Settings
” window opens
(Figure
10).
Click the “
2.
Read Settings from file
” button. Then, select your previ-
ously saved
data restore
file.
Click the “
3.
Write Data to Sensor
” button, Click “
Yes
”, Click “
OK
” and
“
Close
”.
Click the “
4.
Factory
” tab. The Factory Tab window opens. Type the
Gradient value
you recorded from the
Gradient Off Tag
(
Figure 11
).
Data from Device window; Backup and restore settings.
Figure 10.
Factory window; Entering the new gradient value and reading
Figure 11.
new settings
Click “
5.
Read
”.
Click the “
6.
Volume
” tab. The Volume tab window opens. Click ”
Strap
table
”.
Click “
7.
Read from Table
”. Then, click “Write data to sensor”.
You will be prompted for a password. The password is “
8.
becareful
”.