M-Series Model MG Digital Gauge, Transmitter Electronics -  Replacement Guide 551104 A

Installing current parameters

5

Setting new gradient value and board parameters

Your new boards must be programmed before use. Follow the steps 

below to program your new boards:

Launch your setup software and connect to the transmitter. Click 

1. 

the “

Data from Device

” tab, then click the “

Backup/Restore

” button. 

The “

Backup and Restore Device Settings

” window opens 

(Figure 

10).

Click the “

2. 

Read Settings from file

” button. Then, select your previ-

ously saved 

data restore

 file.

Click the “

3. 

Write Data to Sensor

” button, Click “

Yes

”, Click “

OK

” and 

Close

”.

Click the “

4. 

Factory

” tab. The Factory Tab window opens. Type the 

Gradient value

 you recorded from the 

Gradient Off Tag

 (

Figure 11

).

Data from Device window;  Backup and restore settings.

Figure 10. 

Factory  window;  Entering  the  new  gradient  value  and  reading  

Figure 11. 

 

new settings

Click “

5. 

Read

”.

Click the “

6. 

Volume

” tab. The Volume tab window opens. Click ”

Strap 

table

”.

Click “

7. 

Read from Table

”. Then, click “Write data to sensor”.

You will be prompted for a password. The password is “

8. 

becareful

”.