Before you begin

Before replacing your digital transmitter electronics, make sure that it 

is programmed with the transmitter data restore file you created after 

initial installation and calibration of your transmitter.

2

Note: 

 

 

 

 

 

   

Each transmitter requires its own restore file.

To  install  the  transmitter  setup  software,  go  to  the  section  titled 

•  

Software installation and configuration; Modbus and DDA

”.

To create a restore file, go to the section titled “

•  

Create the transmit-

ter data restore file

”. 

If your transmitter electronics PCB is already programmed and you 

•  

have  a  data  restore  file,  go  to  the  section  titled  “

Replacing  your 

transmitter electronics PCB

”.

Create the transmitter data restore file

Obtain the setup software that shipped with your transmitter or go to 

www.mtssensors.com

 and download the latest Modbus or DDA setup 

software from the MTS software Vault.  You will be asked to register 

for a login and password to enter the Vault portal. From the Vault, do 

the following:

For Modbus configuration, download 

Modbus_*.zip

For DDA configuration, download 

DDA_*.zip

Perform the following steps to extract and install the setup software:

Locate and double-click the appropriate 

1. 

protocol_*

.zip file. Extract 

the program folder into C:\ProgramFiles\ and double-click the *.exe 
file to install the program.

Create shortcut and Enable factory mode:

2. 

 Go to C:\ProgramFiles\MTS 

Sensors\ and right click the program configuration folder. Drag the 
folder to your desktop.
Right click the Configuration folder shortcut icon, select “

3. 

Proper-

ties

”. The shortcut properties dialog box opens (

Figure 1

).

Software installation and configuration; Modbus and DDA

Perform the following steps to create a transmitter data restore file:

Click  the  program  file  icon  to  launch  the  configuration  setup  

1. 

software. The setup “Configuration” window opens (

Figure 2

).

Connect  the  transmitter.  If  the  transmitter  is  connected  proper-

2. 

ly, you will see five new tab selections at the top of the window.  
perform the following:

Click the “

a. 

Data from Device

” tab, a new window opens.

Click the “

b. 

Backup/Restore

” button. a new window opens.

Click the “

c. 

Get Data from Sensor

” button, then select the 

       “

Save settings to file

” button.

Type in a  filename such as 

d. 

ModbusRestore

 or 

DDARestore 

      

and path that you can easily locate. then, click “

Save

Note: 

 

 

 

 

 

   

Each  transmitter  requires  its  own  restore  file.  Your  transmitter  must  be 
installed properly and calibrated before you create the data restore file.

Shortcut creation and Target entry

Figure 1. 

Shortcut creation and Target entry

Figure 2. 

In the “

Target

” entry box at the end of the path string, type 

Spacebar/F:1

The string should resemble one of the following:   

 

“C:\ProgramFiles\MTSSensors\ModbusConfig\ModbusConfig.exe”/F:1
“C:\ProgramFiles\MTSSensors\DDAConfig\DDAConfig.exe”/F:1

Continue with “

4. 

Create the transmitter data restore file

”.