Output Operation Procedure
Function
Operation
%
Operation
Display
Setting Range
DCI
20mA
20 mA
%
00.000 mA
-025.00 mA
%
00.000
~
22.000 mA
-025.00
~
112.50 mA
%
Current Output
1) Insert one end of the test leads to the + mA –
output jack
(
OUTPUT
)
of the Instrument and
connect the other end with the input of the
user’s Instrument as shown in Figure 5-1:
M A X
3 0 V
3 0 V
M A X
M A X
3 0 V
IN P U T
O U T P U T
C O M
m A
V
+ L O O P
X M T
+ m A -
+
Figure 5-1
2) Press the key
〔
OUTPUT/IN
〕
, and the
OUTPUT appears in the display. It denotes
that the Instrument is in an output state;
3) Press the key
〔
mA%/V
〕
, to select the output
to be set in ‘mA’or %, and then the unit mA or
mA% appears, in which 0% denotes 4mA;
100% denotes 20mA;
4) Press the key
〔
〕
/
〔
〕
to select the set
digits of the output;
5) Press the key
〔
〕
/
〔
〕
to change the
value of the set digits. The value can do carry
or number decrement automatically. Hold the
pressed key in one second and the numerical
value will keep varying.