Appendix
37
Display reading
not stable
Excessive vibration
Increase filtering or increase ‘d’ in Cal
Excessive side
loading
Improve load train symmetry
Load cell faulty
Check LC connections
Display toggles
between “Error”
and “Load”
Load exceeds
capacity
Reduce tension immediately
Faulty Load cell or
wiring
Check LC and LC wiring
Display toggles
between “Error”
and “buttn”
A key is stuck or is
being held down
Check switches for damage
Weight is on the
Scale/Dyna-Link,
RF Remote
Display does not
match
Units not mated
See setting the RF Network address
procedures
Lo Batt is blinking
Battery is low
Recharge Battery
Unit turns on,
then immediately
off
Battery is low
Recharge Battery
Load will not zero
System not stable
Wait for stable symbol
to turn on
Increase filtering for more stability
Zero out of range
Zero range might be limited. Reduce the
tension or use Tare instead
Load will not Tare
or Total
System not Stable
Wait for the stable symbol
to turn
on, or if in a mechanically noisy crane,
increase the filtering or reduce the size of
the Scale increment “d.” It is also possible
to increase the motion window. Contact
MSI if you have a problem getting the
8000 to zero, tare, or total due to stability
issues
Set point lights
blink
Set point is enabled
and the trigger point
has been reached
Disable set points if they are not needed
Manual Total does
not work
A function key is not
set to “total”
Set up Func1 or Func2 for “total”
Tension must be
stable
Increase filtering for more stability
Problem
Possible Cause
Solution
Table 6-1. Troubleshooting