ML41biological microscope user manual
12
Website
2.2 Installation steps:
2.2.1 Install the stage
(1)Completely loosen the fixing screws
①
on the stage.
(2)Carefully place the two V-shaped buckles on the bottom of the
stage away from the center of the bracket circular hole, and fit them
into the V-shaped circular grooves
②
on the bracket, and then tighten
the fixing screws
①
. (Picture 3)
2.2.2 Install the condenser
(1)Turn coarse hand wheel
①
to raise the stage to the highest position.
(2)Turn condenser lift hand wheel
②
to lower the condenser bracket to
the lowest position.
(3)Completely loosen the condenser fixing screws
③
.
(4)Roll out the front lens group of the condenser, with the numerical
scale facing forward, and align the set screw on the back of the
condenser with the slot on the condenser holder, and push the
condenser into the innermost part of the condenser holder. (Picture 4)
(5)Tighten the condenser fixing screw
③
, then turn the condenser
lifting hand wheel
②
to raise the condenser to the highest position.
2.2.3 Install the objectives
Turn the coarse hand wheel to lower the stage, and then install the
objectives on the nosepiece in a counterclockwise direction from low
to high magnification. (Picture 5)
2.2.4 Install binocular head
(1)Completely loosen the observation base locking screw
①
.
(2)Move the dovetail interface at the bottom of the observation base to
the right from the center of the round hole on the main frame, tilt it
slightly to the left, insert the circular hole along the hole, make the
double cylinder of the observation base face forward, and tighten the
locking screw
①
. (Picture 6)