PrimaX® IR
40
Approvals
GB
Junction Box Typ S47K and X Series Aluminium Junction Box
Special conditions:
The connection cable has to be appropriate for the +80°C ambient temperature range. For group III
application, the apparatus has to be installed in such a way that process-related electrostatic charges,
e.g. caused by media passing by, can be excluded
Manufacturer:
MSA Europe GmbH,
Schlüsselstr.12,
CH-8645 Rapperswil-Jona
Product:
Junction Box Typ S47K and X Series Aluminium
Junction Box
EC-Type Examination Certificate:
BVS 10 ATEX E 066X
Standards:
EN 60079-0:2012 + A11:2013, EN 60079-1:2014,
EN 60079-7:2007, EN 60079-31:2014
Performance:
See MSA Controller together with sensor series 47K
Marking:
Cable gland: NPT ¾” or M25 x 1,5
II 2G Ex db IIC T4/T6 Gb
II 2G Ex e IIC T4/T6 Gb
II 2D Ex tb IIIB T85°C / T135°C Db IP6X
Cable gland type HSK-M-Ex-d
T4
-50°C
≤
Ta
≤
+80°C
T6
-50°C
≤
Ta
≤
+40°C
T85°C
-50°C
≤
Ta
≤
+40°C
T135°C
-50°C
≤
Ta
≤
+80°C
Cable gland
type 8161/5-M25-17
T4
-40°C
≤
Ta
≤
+70°C
T6
-40°C
≤
Ta
≤
+40°C
T85°C
-40°C
≤
Ta
≤
+40°C
T135°C
-40°C
≤
Ta
≤
+80°C
Junction Box Typ S47K with push
button M25 x 1,5
II 2G Ex e IIC T4/T6 Gb
II 2D Ex tb IIIB T85°C / T135°C Db IP6X
T4
-20°C
≤
Ta
≤
+50°C
T6
-20°C
≤
Ta
≤
+40°C
T85°C
-20°C
≤
Ta
≤
+40°C
T135°C
-20°C
≤
Ta
≤
+50°C
Quality Assurance Notification:
0158
Year of Manufacture:
see Label
Serial Nr.:
see Label