PrimaX® IR
31
Maintenance
GB
Fig. 21
Environmental Guard Cleaning
5.6
Calibration Cap Cleaning Procedure
In general, the calibration cap should not require any maintenance. However, if it is used in extremely
dirty or solvent based environments, it may be necessary to clean the light sensor if the calibration cap
performance is not adequate.
(8)
Use a clean, dry applicator and isopropyl alcohol to gently swab the light sensor, as shown in
Fig. 22.
CAUTION!
Never immerse the calibration cap in water or damage will occur.
Fig. 22
Calibration cap diode cleaning