KWIK DRAW UNIPHOS 
Airtester HP 

(Multiport Sampling)

for compressed air according to EN 12021

Instruction for use

1.  Use

The  compressed  air  monitoring  set  “KWIK  DRAW
UNIPHOS Airtester HP” (HP = high pressure) is used 
to determine the content of carbon monoxide, carbon
dioxide,  oil  and  water  vapour  in  respirable  air  from
compressors and compressed gas cylinders.         

   .

According  to  EN  Standard  12021  air  for  respiratory
equipment  must  meet  the  following  quality
requirements: 

CO content max. 15 ppm
CO  content max. 500 ppm

2

Free  of  odor  and  taste,  i.e.  practically  free  of
critical mineral oil content (from experience the oil
content is less than 0.5 mg/m³)
Water content in compressed air cylinders:
max. 50 mg/m³ at a charging pressure of 200 bar
max. 35 mg/m³ at a charging pressure of 300 bar
max. 25 mg/m³ of the air from the compressor

-
-
-

-

The  requirements  concerning  the  maximum  water
vapour content refer to the type of equipment used
(200 or 300 bar) and, therefore, apply also to each
pressure  level  of  the  compressed  gas  cylinders
below the maximum values given above.     

          .

All  maximum  values  are  based  on  air  not  under
pressure.

2. Measuring Principle

3. Design

The case contents all parts required for measuring
(see fig. 1): 

-

The standard case contents one package of each of

- 3 -

Fig. 1 KWIKDRAW UNIPHOS  Airtester HP 

4. Technical Data

Measuring ranges of the Detector Tubes:

Carbon monoxide detector tubes

5  to  70  ppm  carbon  monoxide  at  a  flow  rate  of 
0.3 l/min in a measuring time of 5 minutes. (required 
sampling volume 1.5 l).

See also specific instructions for use of the KWIK 
DRAW / MSA AUER detector tubes CO-HP.

Carbon dioxide detector tubes

100  to  2000  ppm  carbon  dioxide  at  a  flow  rate  of 
0.3 l/min in a measuring time of 5 minutes. (required 
sampling volume 1.5 l).

See   also   specific   instructions    for   use   of   the 
KWIK DRAW / MSA AUER detector tubes CO2-HP.

Oil detector tubes

This type of tube indicates semi-quantitatively mineral 
and synthetic Oils.

Check  for  oil  contents  (vapor  +  mist)  at  a  flow  rate  of 
2.5 l/min in a measuring time of 24 minutes. (required 
sampling volume 60 l). The content of oil is determined 
as  per  the  instruction  given  to    use    the  KWIK 
DRAW / MSA detector  tube Oil-HP Synth. 

the four types detector tubes. In addition, the case 
has  space  for  storing  four  additional  spare
packages.

air
300 + 30 bar
 0.3 , 1.0  and 2.5 l/min. 
 (The detector tubes are to be 
fitted in an appropriate port of 
manifold to pass required 
sample air.)  
5 kg
h= 120; w= 320; l= 420 mm

Operation medium:
Maximum pressure:
Flow:

Weight (with case): 
Dimensions: 

- 6 -

6. Maintenance 

6.1 General

In  view  of  the  significance  of  the  accuracy  of 
measuring  results,  the  test  device  must  be  treated 
with care. Dirt must be avoided at all time, especially 
at the high-pressure connection.

Flushing  and  cleaning  of  the  test  device  must  be
made  as  described  in  para.  5.2. After  measuring  is 
completed, the test device shell is stored in the case.
The high pressure connection should be closed with 
a protective cap.

Function checks should be made at regular intervals 
and if erroneous measurements are suspected.

6.2 Disorders

If  the  hand  wheel  of  the  test  device  cannot  be 
loosened while it is connected to a compressed gas 
cylinder or compressor, check if connection is under 
pressure.  In  this  case  close  the  cylinder  valve  or 
compressor valve and release air from test device.

Leaks  at  the  connection  to  the  cylinder  or 
compressor could be due to a defect gasket. In this 
case replace the gasket.

If the high pressure gauge or the detector tube holder 
is defective, they must be replaced.

In  case  of  other  disorders  the  device  should  be 
returned for repair to  UNIPHOS or any authorized 
service station.

7. Part Numbers

KWIK DRAW UNIPHOS Airtester HP

(Compressed air monitoring set)

Complete with case containing:
Test  device  consisting  of pressure  reducer with 
high pressure gauge, manifold, flow meter detector 
tube holder.

Adapter for 200 and 300 bar one each for connecting 
to compressor charging panel

Timer

Device for breaking off detector tube tips,

Detector   tubes  for  measuring  carbon  monoxide, 
carbon dioxide, water vapor, oil - one package each.

Accessories and Spare Parts

Detector tubes CO-HP* 

Detector tubes CO -HP* 

2

Detector tubes H O-HP* 

2

* = Package of 10 pieces

D5085847

D5085848

D5085849

Detector tubes Oil- HP Synth.*
for testing of mineral and synthetic Oil

  

Pressure reducer with pressure gauge 

Manifold

Gaskets-lifter 

Detector tube holder 

Timer 

Tube tip breaking device 

Case 

Case insert 

Flow meter

10040887

D4074937

CR001444

D4060241

D4074058

D5185020

D5085012

D5185911

D5185931

CR001452

The  pressure  in  the  compressed  gas  cylinders,  or 
in  the  compressor  charging  panel  is  reduced  by  a 
pressure  reducer.  The  air  to  be  sampled  is  taken 
continuously    via  a  regulatory  valve  in  the  manifold. 
There are four outlets for this manifold which release
fixed  flow  rate  of  gas.  The  flow  rate  of  each  port  is 
printed  on  the  body  of  the  manifold,  viz  0.3  l/min, 
0.3  l/min  1.0  l/min  and  2.5  l/min.  The  air  coming  out 
from  these  ports  is  taken  through  detector  tube  for  a
specific time. The length of stain in the detector tube 
is  an  index  of  the  concentration  of  the  measured 
substance. This can be read directly from the detector
tube scale.

The test device, consisting of pressure reducer,
Pressure  gauge,  manifold,  flow  meter,  detector
tube  holder,  adapter  for  connection  to  a 
compressor  charging  panel  (  for  charging 
connection 300 or 200 bar), timer, detector tube
tip breaker, tubes (package of 10) for measuring
CO,  CO2,  water  vapor  and  oil  content  (sum  of 
Oil  vapor  and  mist).                              .