Troubleshooting
*Sensor Warning and Sensor Alarm indicate that the device detects a loss of sensor sensitivity during
calibration. In addition to true loss of sensor sensitivity, the device may issue a Sensor Warning or
Sensor Alarm if calibration gas is expired, misapplied or not used during the span calibration. Ensure
the quality of the calibration gas and the calibration gas delivery system. Recalibration may remove the
Sensor Warning or Sensor Alarm indication.
Problem
Description
Reaction
ERROR TEMP
Temperature is below
-40°C (-40 °F) or above 75°C (167 °F).
Return device to normal
temperature range and recalibrate.
Contact MSA
ERROR EE
EEPROM Memory error
Contact MSA
ERROR PRG
Flash Memory error
Contact MSA
ERROR RAM
RAM Memory error
Contact MSA
ERROR BT
Bluetooth error
Contact MSA
ERROR UNK
Unknown error
Contact MSA
LOW BATT
Battery warning repeats every
60 seconds.
Remove from service as soon as
possible and recharge battery
BATT
ALARM
Battery is completely discharged.
Device is no longer sensing gas.
Remove from service and recharge
battery.
ERROR CHARGE
Charge error
Device must be between 10° C and
36° C to charge.
Contact MSA if problem persists
SENSOR ERROR
Missing Sensor
Verify if sensor is properly installed
Device does not
turn on
Low battery
Charge device
Sensor Warning
Sensor is near the end of its life
Sensor Alarm
Sensor has reached the end of its life
and cannot be calibrated. Replace
sensor and recalibrate.
or
(flashing)*
*
Summary of Contents for ALTAIR 4XR
Page 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Page 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Page 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Page 55: ...ALTAIR 4XR 55 Appendix US ...
Page 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Page 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Page 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Page 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Page 60: ...ALTAIR 4XR 60 Appendix US ...
Page 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Page 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Page 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...