ALTAIR
®
4XR
30
Operation
US
Fig. 5
Calibration Setup
(1)
To bypass this setup, press the
or
button; otherwise, continue as follows.
(2)
Press the
q
button to enter the submenu.
The calibration gas concentration of the first sensor is shown.
(3)
Press the
or
button to change the value.
(4)
Press the
q
button to store the value.
Setup screen for the next sensor is displayed.
(5)
Repeat the procedure for all other sensors.
After setting the last sensor, the user is prompted to set CALDUE.
(6)
Press the
or
button to enable or disable CALDUE.
(7)
Confirm with the
q
button.
(8)
If CALDUE is set on, press the
or
button to select the number of days
(9)
Confirm with the
q
button.
(10) Press the
or
button to enable or disable the calibration password (CAL PW) screen.
(11)
Confirm with the
q
button.
(12) After confirmation, continue to Alarm setup.
Alarm Setup
The user can switch all alarms on or off and change the alarm set points for each sensor. For more
information, see the flow charts in chapter 7.8 "Alarm Setup".
See chapter 5.1 "Factory-set Alarm Thresholds and Setpoints" for alarm adjustment limits. The value
of the High Alarm can only be set to a value that is higher than the Low Alarm set point.
Fig. 6
Alarm Setup
(1)
To bypass this setup, press the
or
button; otherwise, continue as follows.
(2)
Press the
q
button to enter the submenu.
Summary of Contents for ALTAIR 4XR
Page 1: ...MSAsafety com Operating Manual ALTAIR 4XR Multigas Detector Order No 10175896 00 ...
Page 52: ...ALTAIR 4XR 52 Appendix US 7 Appendix 7 1 Start Up Sequence Power on ...
Page 53: ...ALTAIR 4XR 53 Appendix US 7 2 Fresh Air Setup FAS ...
Page 55: ...ALTAIR 4XR 55 Appendix US ...
Page 56: ...ALTAIR 4XR 56 Appendix US 7 4 Bump Test ...
Page 57: ...ALTAIR 4XR 57 Appendix US 7 5 Options Setup ...
Page 58: ...ALTAIR 4XR 58 Appendix US 7 6 Sensor Setup ...
Page 59: ...ALTAIR 4XR 59 Appendix US 7 7 Calibrations ...
Page 60: ...ALTAIR 4XR 60 Appendix US ...
Page 61: ...ALTAIR 4XR 61 Appendix US 7 8 Alarm Setup ...
Page 62: ...ALTAIR 4XR 62 Appendix US 7 9 Time and Date Setup ...
Page 63: ...sales norrscope com ALTAIR 4XR 63 Appendix US 7 10 Bluetooth Setup ...