Operation
MSA
ALTAIR 4X
GB
35
(5) Press the
or
button to change the value.
(6) Press the
ON/OFF
button to store the value.
HIGH ALARM settings for the first sensor display.
(7) Press the
or
button to change the value.
(8) Press the
ON/OFF
button to store the value.
STEL ALARM settings [for toxic sensors only] display.
(9) Press the
or
button to change the value.
(10) Press the
ON/OFF
button to store the value.
TWA ALARM settings [for toxic sensor only] for display.
(11) Press the
or
button to change the value.
(12) Press the
ON/OFF
button to store the value.
(13) Repeat the procedure for all other sensors.
(14) After setting the last sensor, continue to Time and Date setup.
60% L.E.L. or 3.0% volume of methane is the maximum High Alarm set point that
can be programmed.
The alarm can be silenced momentarily by pressing the
button. However, if the
gas concentration causing the alarm is still present, the device will go back into
alarm.