USER MANUAL
OPTIMA 7
MRU GmbH, D-74172 Neckarsulm
54 / 78
With this function you can Import Sites which have been created on a
computer or another Analyzer.
The File name must have the name “anlagen.csv“(anlagen =
German for
sites). The file has no column heading that means that the first line al-
ready has user data. Each line has a minimum of 9 columns (with 8
semi-colons) and the first field in the line will be the site number. All
data will be imported as long a site number is available. Per field a maxi-
mum of 24 characters will be imported, too long words will be cut off.
Example file with 8 valid sites (4 with 9 lines and 4 with less lines):
A1-Z1;A1-Z2;A1-Z3;A1-Z4;A1-Z5;A1-Z6;A1-Z7;A1-Z8;A1-Z9
A2-Z1;A2-Z2;A2-Z3;A2-Z4;A2-Z5;A2-Z6;A2-Z7;A2-Z8;A2-Z9
A3-Z1;A3-Z2;A3-Z3;A3-Z4;A3-Z5;A3-Z6;A3-Z7;A3-Z8;A3-Z9
A4-Z1;A4-Z2;A4-Z3;A4-Z4;A4-Z5;A4-Z6;A4-Z7;A4-Z8;A4-Z9
A5-Z1;A5-Z2;A5-Z3;A5-Z4;;;;;
A6-Z1;A6-Z2;;A6-Z4;;;;;
A7-Z1;;;A7-Z4;;;;;
A8-Z1;;;;;;;;
Example file with 2 invalid sites (1 with not enough fields and 1 with
missing site number):
A1-Z1;A1-Z2
;A1-Z2;A1-Z3;A1-Z4;A1-Z5;A1-Z6;A1-Z7;A1-Z8;A1-Z9
NOTE
While importing data from the SD Card to the analyzer
there is no check for double site numbers (Line 1), nei-
ther inside of the file that is imported nor between the
file and the sites already inside the analyzer. The ana-
lyzer can easily handle double site numbers but you
could face problems with double site numbers when ex-
porting them again to a computer program (see also Ex-
port of Measurements).
However, the analyzer marks the files that have been im-
ported successfully. If you try to import a file with the
same analyzer that is already in the analyzer you will get a
red information screen.