USER MANUAL
OPTIMA 7
MRU GmbH, D-74172 Neckarsulm
14 / 78
The Condensate separator (water trap)
•
CAUTION
Condensate is acidic
The condensate from the container can be slight acidic
and can cause chemical burn.
►
Immediately clean with plenty of water once you have
encountered acid
During the measurement accumulating condensate is collected in the
condensate separator
Remove the condensate separator by pulling it towards you (1) out of
the groove of the OPTIMA 7 housing, then pull it downwards (2).
Condensate vessel and plugs are screwed together. To change the filter
and dry and clean the condensate separator this can be easily disassem-
bled and screwed together again.
After cleaning, a leak test must be carried out.
See also chapter 9.4 Performing leak test, Page 61.
In the condensate separator of the OPTIMA7 an exhaust gas filter is at-
tached. This filter out the first coarse dirt. This pocket star filter is washa-
ble and can be used 4-5 times afterwards.
Frequent measurements result in a high degree of contamination which
requires regular replacement of this filter.