USER MANUAL
MGAprime
90 / 126
MRU GmbH, D-74172 Neckarsulm
9.4.
Internal log settings
The analyser stores internal parameters at regular intervals to allow for an
optimized support from experienced service staff. These files can be copied
to a connected USB stick and sent via email if required to do so. The data will
be anonymized. It is not possible to identify sites or
measuring points.
Copying all log data
►
Insert a USB-stick into the analyser.
►
Open the menu Service.
See also chapter 9.1 Open menu Service, S.88.
►
Press Internal log settings.
A window appears.
►
Press „SERVICE:COPY INTERNAL LOGG TO USB STICK“.
A window appears.
The copy process is started.
A Symbol appears in the Display
►
Press OK.
The directory [serial number] Logg was created on the USB-Stick.
The log files are saved in the directory [serial number].
After copying, the display shows "File(s) successfully copied".
►
Touch OK.
NOTE
If the analyser is operated for a long time, the copying pro-
cess can take a long time.