Hand Wheel Vertical Autoclave

User’s Manual

STE-V60/80/100

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 

3, Hagavish st. Israel 58817  Tel: 972 3 5595252, Fax: 972 3 5594529   

[email protected]

 

MRC.

4

.1

7

 

Summary of Contents for STE-V100

Page 1: ...Hand Wheel Vertical Autoclave User s Manual STE V60 80 100 PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 4 17 ...

Page 2: ...erilization 15 large package of difficult sterilization should be placed in the upper the smaller packet should be placed on the lower deck 15 sterilizer loading volume shall not exceed 80 15 only liquid bottles and tubes with the heat load the load should not exceed the amount of container volume 15 IV Warranty 20 Figure 1 Electrical Schematic Diagram 21 Figure 2 PID Pipeline Schematic 21 Figure ...

Page 3: ...ttention to personnel safety 2 2 Equipment inspection After opening the packing box please carefully check whether the equipment and parts are in good condition if there are any damage or loss please kindly make record and contact our company After unpacking the equipment firstly check the model and product name on the product nameplate whether compliance with the order Product nameplate is at the...

Page 4: ...chine feet make them parallel to the ground to ensure stable reliable The distance between back and sides of the sterilizer and other objects at least 250 mm To ensure the ventilation is well In addition to keep a safe distance from other objects the existence of other objects should not affect the equipment operation When there is a failure in the equipment operation you should be able to cut off...

Page 5: ...sure to lay a ground wire 3 Water source required Devices do not need to connect the water you need to add water to sterilizer water tank or sterilization chamber manually You are advised to use soft water or pure water Because if use water which is not suitable may shorten the service life of equipment cause unnecessary trouble Water quality must meet the following requirements Electrical conduct...

Page 6: ...age period kept complete instructions for use Ensure that all updates received can be preserved in the specification In the device using the site or the use of units of change we must ensure that the instruction for use as part of the overall transfer or transfer equipment 2 Equipment marking instructions To show should pay great attention The sign of opening the door The sign of high water level ...

Page 7: ...ei card characters 2014 504th Design service life 5 years Minimum equipment sales unit 1 Attention 1 Please open the door after discharging all the steam Do not close to the sterilizer at the same time 2 The chamber is still keep at high temperature after the sterilizing please pay attention to insulate the chamber in case of scald 3 Monitoring method Methods to monitor sterilization effect includ...

Page 8: ...n or close the door of autoclave 2 Door cover Heat insulation and protect the operator 3 Water mouth deionized water is injected from water month to the water tank 4 Water level indicator tube Indicates water level in the water tank 5 Drainage mouth Drain the dirty water after cleaning the sterilization 6 Universal wheel Make the sterilization move smoothly 7 The relief valve Relief the pressure t...

Page 9: ... at which pulse exhaust starts The temp at which pulse exhaust finishes Steril izing temp Sterilizing time min Drying time min Heat retaining time min 1 Apparatus 1 Sterilizing temp 2 5 The temp at which pulse exhaust starts 5 1 34 4 5 0 2 Apparatus bag 2 Sterilizing temp 2 5 The temp at which pulse exhaust starts 5 1 34 8 10 0 3 Cloth 3 Sterilizing temp 2 5 The temp at which pulse exhaust starts ...

Page 10: ...ts Add water Temp rising Sterilizing Air discharge Drying Finish Alarm Liquid and Door and Temp display and Time display Operating procedure Plug in switch on the air switch on the left and then turn on the rocker switch on the panel The panel lights on the autoclave stands by Temp diaplay shows the word Prog Time display shows number X X is the number from 1 to 7 which refers to the chosen progra...

Page 11: ...re button both for 5 seconds to enter the menu Display iwti represents the time to add water unit minute when press the Start End the button shows the value and flashes Through to change the value press Start End to confirm Must be 0 displayed iwti press the EATE displayed and press the tErF displayed EATE represents the cooling end temperature press the Start Stop the button displays the value an...

Page 12: ...value and flashes through to change value press Start End to confirm Only custom solid program valid default is 10 After displaying drti press the tEFA displayed if press ESTi after Display tEFA indicates the temperature calibration factor press the Start End then shows the value and flashes through to change value press start end to confirm and users are strictly prohibited to modify When display...

Page 13: ...Run Sterilizer operating procedures include sterilization preparation sterilization items loading sterilization sterile unloading procedure 6 1 Sterilization preparation 1 Cleaning article should be thoroughly cleaned before sterilization avoiding the presence of blood and other impurities as these substances would harm sterilizing articles and sterilizer After washing items should be timely dried...

Page 14: ...leaching textiles cloth seamless addition to the four sides hot wash before first use should skim to pulp to color should use records 6 2 items loading sterilized items please follow the requirements for loading When the items put in up and down should be a certain distance from each other items can not be posted near the door and walls to prevent inhalation of more condensate similar materials an...

Page 15: ...e that the sterilizer is in good working condition and minimize the number of failures therefore the operation described in this chapter must be followed Before starting maintenance make sure the device is powered down At the same time no pressure in sterile containers After the end of the working day with a soft cloth or a piece of gauze to wipe the door gasket Remove the basket With gauze with a...

Page 16: ...he maximum and minimum water 7 2 Exhaust steam valve cleaning steps Before you begin you should cut off the power supply and ensure that no pressure inside the sterilization chamber 1 Remove the tank lid 2 by stretching before and after steam escape valve flush valve orifice steam escape 3 If it is necessary to replace the steam escape valve before replacing to wait for equipment cooling the press...

Page 17: ...able value the thermostat automatically disconnect the heater power When the pot temperature dropped below the allowable value the thermostat automatically turns on 7 6 How to improve the operating temperature of the thermostat This operation is limited to professionals Use a screwdriver to rotate the center screw clockwise slightly so that the temperature rise 7 7 replace the heater step Before t...

Page 18: ... slip on the electrical magnetic valve with a screwdriver 3 lift the electromagnetic valve coil 4 open the valve body with a spanner 5 clean sundries on the valve center with clean water 6 re install the electromagnetic valve Third Common fault and solution 1 This Instruction tries to provide the maintenance method for the known fault the below are some common faults Phenomena Possible reasons Cor...

Page 19: ...et temperature of the boiling point If not the reason of the altitude please contact us or our distributor Safety Valve opens 1 Whether the pressure is too high 2 Whether the safety valve breaks down 1 Adjust temperature departure 2 Adjust and change safety valve Steam leaking 1 W hether door rubber gasket become hard and aged 2 W hether the door strip cracks 3 Whether door rubber gasket falls off...

Page 20: ...noid valve Err 064 The water shortage in the pot caused heating pipe burned Float failure after the heating pipe cool down add enough water to the pot Err 128 After starting sterilization the pan is opened excessive pressure the pot shifts pot test failure IV Warranty To our distinguished guests Any product has the possibility of failure please monitor the work of the equipment in real time during...

Page 21: ...Figure 1 Electrical Schematic Diagram Figure 2 PID Pipeline Schematic ...

Page 22: ...erial Number Description Quantity Notes 1 Equipment 1 2 Operating instruction 1 3 Pressure vessel Quality certificates 1 4 The warranty card 1 5 Conformity certificate 1 6 Mesh basket 2 7 Water fender 1 8 Tank lid 1 9 Seal ring 1 10 Drain pipe 1 ...