Caution: When you press the "START” button but the door have not to be
closed. You will see the
blinks on the screen, You can not
start a cycle until you close the door to the max. position and
press the "Start" button again.
9
5.4.2 Start the sterilization program.
Press START button the machine will begin the cycle automatically.
It will take 30-75 minutes. (See Appendix 2)
total time or
time count down
of holding time
and end time
total time or
time count down
of holding time
and end time
After a cycle is completed, the printer will start work and print the report
of the sterilization cycle data.(if you connect the printer)
After the pressure is 0, you may open the door, and take out the sterilized
instruments.
5.4.3 Sterilization cycle completion
Always use the tray handle to load or unload the tray in order to
avoid scald.
1 0
If you need the instrument urgently. You may set the drying time
to 0. It will skip the drying stage.
Or you may holding the START button for 3 seconds after finish
the holding time and the pressure is 0.
The program will go directly to the last step and skip the drying
stage. After one minute later it will show END.
(CLASS B)
(CLASS B)
(CLASS N)
(CLASS N)
Instructions manual