Distilled water tank is lacking of water.
Distilled water tank is full
Used water tank is full.
Door locked
information output to USB port
Printer connect
Notice: Before using the sterilizer at the first start-up or any time the
icon
blink, it is necessary to fill the distilled water tank
with distilled water.
5.1 Ready
5.1.1 Open the door and take out all the instrument tray and other
accessories inside, unpack and clean them.
5.1.2 Connect the power, and connect the printer (See 6.5)
5.1.3 Power on
The switch is located underneath the control panel
on the front side of the machine .
After switching on, the machine turns on the LCD .
Then it will show the door position, water level,
working program, date, time and etc. .
The machine will alert by beep sounds if there are
problems.
5 Operation
Program
Alarm code
process curve
T
emperature
Pressure
Setting
Time
Date
5
5.1.4 Filling the distilled water
Open the top lid, and fill the tank with distilled water by cup or tank.
When you hear a beep signal, it means the water level exceeds the
max. level. The
will be displayed. Please stop filling immediately.
5.2 Preparing the material to be sterilized
To get the better effectiveness of the sterilization process and to preserve
the material in time, follow the indications below reported.
* Arrange the tools of different metal (stainless steel, moderate steel,
aluminum, etc.) on different trays or however well separate between them;
* In case of not stainless steel tools, interpose a sterilization paper napkin or
muslin cloth between tray and tool, avoiding direct contacts between the
two different materials;
* Verify all the tools are sterilized in open position;
* Arrange the containers (glasses, cups, test-tubes, etc.) on one side or
inverted position, avoiding possible water stagnation;
* Don't overload the trays over the stated limit (see Appendix 1).
* Don't stack the trays one above the other or put them in direct contact with
the walls of the sterilization chamber.
* Always use the instrument tray handle.
* Wrap the tools one by one or, if more tools have to be set in the same wrap,
verify that they are of the same metal;
*Seal the wrap with sterilization adhesive ribbon or by a thermal sealer.
* Don't use metallic clips, pins or other, as this jeopardizes the maintenance
of the sterility;
6
The water level should not
exceeded this port.