The 16 digit data stream will be displayed in the
following format :
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
Each digit indicate the following status :
D0
End Word
D1 & D8
Display reading, D1 = LSD, D8 = MSD
For example :
If the display reading is 1234, then D8 to D1 is :
00001234
D9
Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D10
Polarity
0 = Positive
1 = Negative
D11 & D12 Annunciator for Display
Bar = 22
mm/Hg = 78
inch/HF2
= 25
Psi = 23
inch/Hg = 80
ATP = 26
Kg/cm
= 77
meter/HF2
= 79
D13
1
D14
4
D15
Start Word
9600, N, 8, 1