2) Connect the pressure regulator to the nitrogen bottle
and the reference gauge to the pressure regulator.
The pressure regulator should can generate the full scale
pressure value.
For example, input a standard 9.00 pressure value to the
the sensor, the display show 8.98 Bar.
3) At the same time hold the following two button continuously .
" Hold Button " ( 4-3, Fig. 1 )
" Max./Min. Button " ( 4-4, Fig. 1 )
The display show the measuring value on both up & down
display.
For example :
Bar
.
8.98
4) Still hold the " Hold Button " & the " Max./Min. Button ",
at same time.
*
Push the " Sensor Type Button " ( 4-7, Fig. 1 ) once
will add one count to the down display value.
*
Push the " Unit Button " ( 4-5, Fig. 1 ) once will decrease
one count to the down display value.
13