6. MEASURING PROCEDURE
1) Plug in the " Plug of Pressure Sensor " ( 4-13, Fig. 1 ) to
meter's " Sensor Input Socket " ( 4-9, Fig. 1 )
2) Power on the meter by push the " Power Button " ( 4-2,
Fig. 1 )
3) Push the " Sensor Type Button " ( 4-7, Fig 1 ) to check if
the meter's sensor type is same as the external pressure sensor
4) Push the " Unit Button " ( 4-5, Fig. 1 ) to select the unit (
Bar, Psi, Kg/cm 2 mm/Hg, inch/Hg, meter/HF2 ,
inch/HF2 , Atmosphere ).
5) Zero the meter by push the " Zero Button " ( 4-6, Fig. 1 ),
the display reading will change to zero value.
6)
Connect the " Port Connector of Pressure Sensor "
( 4-11, Fig. 1 ) to the installation that intend to measure
the pressure value that into the sensor.
7) Apply the pressure, pressure meter will show the pressure
value.
8) Data Hold
During the measurement, pushing the " Data Hold
Button " ( 4-3, Fig. 1 ) will freeze the measured value &
display will indicate " HOLD " symbol.
* Push the "Data Hold Button" again to release the
data hold function.
9)
Data Record ( Maximum, Minimum reading )
* The DATA RECORD function displays the maximum
and minimum readings. To start the DATA RECORD
function, press the " Max./Min. Button " ( 4-4, Fig. 1 )
once. " REC " symbol will appear on the LCD display.
8