INE-

PHSJ-5 pH Meter 

13 

1 .When user calibrate electrode with his own standard buffer solution 

( Non-standard buffer solution), he must know the pH nominal value of this 

non-standard buffer solution under the range of calibrating temperature. 

2. In each measurement, suggest to re-calibrating the electrode before hand. As

soon as you start to calibrate, the previous calibration data will be replaced. 

Before calibration, please prepare 1~5 different kinds of standard buffer solution ( it can be normal 
standard  buffer  solution,  and  can  also  be  user‟s  own  standard  buffer  solution),  leave  them  under 

normal  temperature  for  a  while,  then  set  the  standard  buffer  solution  according  to  the 

CALIBRATION FUNCTION(8). Electrode calibrating will be effective after setting. 

Clean  the  pH  measuring  electrode,  reference  electrode,  temperature  sensor  and  put  them  into  the 

measured standard buffer solution altogether. When the reading is stable, press “ENTER” key. The 
meter  shows “STD  Result”.  After  a  while,  the  meter  will  ask  you  “End  Standardization?”(refer  to 

diagram  12);  if  user  need  to  calibrate  some  other  standard  buffer  solution,  please  press 
“CONTINUE” button, then repeat above procedure till calibration finished. 

Diagram 12 

During the calibration, user can finished the calibration at any time by pressing “

End STD

” button. 

After calibrating the fifth standard buffer solution, the meter will finish the calibrating automatically. 

For the standard buffer solution with the range of pH value overlapped, such as the buffer solution of 

6.865 pH & 7.000pH, suggest to calibrate as follows: 

 

When  user  calibrates  6.865pH  standard  buffer  solution,  set  standard  buffer  solution  group  to

6.865pH only, then arrange calibration. When finished, reset standard buffer solution group to 

7.000pH only, then calibrate. 

 

Adopt manual distinguishing mode to calibrate. Every time, when user calibrate standard buffer

solution, input nominal pH value under present temperature in correspond to present  standard 

buffer solution by manual.   

As to normal standard buffer solution, user can use  automatic distinguishing function  with the 

standard buffer  solution group  set before hand. No need to change  distinguishing mode  (if  not 

able to distinguish, the meter will point out: calibration wrong to ask you to exchange electrode 

Summary of Contents for INE-PHSJ-5

Page 1: ...INE PHSJ 5 pH Meter Instruction Manual 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 12 18...

Page 2: ...eter 1 ADVICE Read this instruction in detail before operating the meter The meter should be re tested by metrological department when it is used over one year And the meter can t be used until it is...

Page 3: ...INE PHSJ 5 pH Meter 2 CONTENT 1 GENERAL 2 SPECIFICATIONS 3 CONSTRUCTION 4 OPERATION 5 MAINTENANCE AND SERVICE 6 THE COMPLETE SETS OF THE METER 7 APPENDIX...

Page 4: ...can be connected with TP 16 printer to print out measuring results or be connected with computer to communicate with outside Protection from electricity cutting off The stored data and parameters wil...

Page 5: ...ng terminal 6 measuring electrode socket 7 temperature sensor socket 8 USB interface 9 RS 232 interface 10 power switch 11 power supply socket with fuse 3 Accessory of Meter 11 Q9 short circuit plug 1...

Page 6: ...ap off the fuse socket put 0 5A 220V fuse into it and then twist the fuse cap on Connect power line with the meter Note In general the fuse has been placed in before ex factory If user matched meter w...

Page 7: ...button SAVE button PRINT button and DEL button to enter corresponding working state directly 4 Set function User should check the system of the meter before using including system time types of temper...

Page 8: ...er to dia4 Diagram 4 Diagram 4 Press MONTH button repeatedly to select month you need press SETUP button to enter or use entering method press button spring out a entering window enter your needed mon...

Page 9: ...TUP button select item Default Temp press ENTER button to enter manual setting temperature Press SETUP button or direct press it once in the area of manual temperature setting spring out an input wind...

Page 10: ...electrode at 0 is not exactly 1000 At this time in order to ensure measuring precision it is neccessary to reset R0 value according to nominal resistance value Press SETUP R0 key and input R0 value o...

Page 11: ...standard buffer solution group 5 2 Setting standard buffer solution group The meter has automatic distinguishing function which can distinguish 10 different standard buffer solutions Multi point not...

Page 12: ...avoid overlapping of pH value user should select corresponding standard buffer solution to calibrate User should delete all the other selection of solution which are not used When selection finished p...

Page 13: ...e slope set as 100 potential value and temperature value In the middle of the screen it shows the current standard buffer solution group set by user At the bottom of the screen it shows the result of...

Page 14: ...iagram 12 if user need to calibrate some other standard buffer solution please press CONTINUE button then repeat above procedure till calibration finished Diagram 12 During the calibration user can fi...

Page 15: ...springs out a window and ask Save current result Press ENTER button to store Press CANCEL button to cancel The meter returns to starting working state 8 Print communication function Only one socket or...

Page 16: ...delete a certain stored data At VIEW Stored Datum mode press DEL button to select Delete current one press ENTER button to delete Press CANCEL button to exit the operation Note You should be very care...

Page 17: ...9 4 060 6 833 55 1 715 3 554 4 075 6 834 60 1 723 3 560 4 091 6 836 70 1 743 3 580 4 126 6 845 80 1 766 3 609 4 164 6 859 90 1 792 3 650 4 205 6 877 95 1 806 3 674 4 227 6 886 CLASS DENANARE 1th GRADE...

Page 18: ...E KHC4H4O6 25 SATURATION 7 3 776 1 GRADE KH2C6H5O7 0 05 11 41 4 008 1 GRADE KHC8H4O4 0 05 10 12 6 865 1 GRADE KH2PO4 0 025 3 387 Na2HPO4 0 025 3 533 7 00 DENANARE KH2PO4 0 02 2 780 Na2HPO4 0 0275 3 88...

Page 19: ...t 2 E 201D PH Combination Electrode 1 3 T 820D Temperature Sensor 1 4 Buffer Solution 4 7 10 5 bags for each 5 Multifunction Electrode Support Stand 1 6 Fuse 0 5A 5 7 AF30 Stylus 1 8 TP 16 Print Conne...

Reviews: