background image

INE-

PHSJ-5 pH Meter 

10 

5.1 View calibrating data 

Viewing previous calibration data

calibration time

calibration temperature and current slope 

value; set your own standard buffer solution. Select “Last STD Datum” in “VIEW” menu, press 
“ENTER” button to enter “

VIEW Last STD Datum” mode

. Please refer to Diagram 8: 

 

The 

last 

calibrating 

data 

is

displaying  on  the  top  of  screen, 

including 

calibration 

time

calibrating  point 

calibrating  data 

of  corresponding  standard  buffer 

solution  and  current  percentage  of 

slope value;   

 

The  standard  buffer  solution  group

is  displaying  at  the  bottom  of  the 

screen;   

 

The  “<= =>”(moving)

“STD" and

“SETUP”  button  is  on  the  right  of 

screen. 

When last calibration is multi-calibrating, user can press “<= =>”(moving) button to display detailed 

data of each standard buffer solution under each point. 
At  this  time,  press  “STD”  button  to  calibrate  electrode  slope  directly.  (see  calibration  item  for 

specific procedure) 

 

Press “X”(exit) to exit viewing mode, the meter is back to starting working state.

 

Press “SETUP” button to set new standard buffer solution group.

5.2 Setting standard buffer solution group 
     

The  meter  has  automatic  distinguishing  function,  which  can  distinguish  10  different  standard 

buffer solutions. Multi-point (not exceed 5 point) calibration is applicable. In order to ensure the 

accuracy of the measurement, user should set up corresponding standard buffer solution group, 

to avoid range of pH value overlapping. Otherwise the meter will not distinguish different kinds 

of standard buffer solution, which will influence the result of calibration or even give you wrong 

information.   

For  example,  when  user  adopt  4.008pH

9.180pH  standard  buffer  solution  to  calibrate  slope  of 

electrode, you can set standard buffer solution group as 4.008pH & 9.180pH. In actual calibration, 

the meter is able to distinguish these two kinds of standard buffer solution automatically.   

 

Press  “SETUP”  button  to  enter  setting  current  standard  buffer  solution  mode.  Please  refer  to

diagram 9. 

 

10  icons  in  the  “Set  BUFFER  GROUP”  window  is  corresponding    to  10  kinds  of  standard

buffer solutions, they are 1.679pH

3.557pH

3.776pH

4.008pH

6.865pH

7.000pH

7.413pH

9.180pH

10.012pH

12.454pH. The nominal pH value of the standard buffer solution is shown 

under each icon.   

Summary of Contents for INE-PHSJ-5

Page 1: ...INE PHSJ 5 pH Meter Instruction Manual 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 12 18...

Page 2: ...eter 1 ADVICE Read this instruction in detail before operating the meter The meter should be re tested by metrological department when it is used over one year And the meter can t be used until it is...

Page 3: ...INE PHSJ 5 pH Meter 2 CONTENT 1 GENERAL 2 SPECIFICATIONS 3 CONSTRUCTION 4 OPERATION 5 MAINTENANCE AND SERVICE 6 THE COMPLETE SETS OF THE METER 7 APPENDIX...

Page 4: ...can be connected with TP 16 printer to print out measuring results or be connected with computer to communicate with outside Protection from electricity cutting off The stored data and parameters wil...

Page 5: ...ng terminal 6 measuring electrode socket 7 temperature sensor socket 8 USB interface 9 RS 232 interface 10 power switch 11 power supply socket with fuse 3 Accessory of Meter 11 Q9 short circuit plug 1...

Page 6: ...ap off the fuse socket put 0 5A 220V fuse into it and then twist the fuse cap on Connect power line with the meter Note In general the fuse has been placed in before ex factory If user matched meter w...

Page 7: ...button SAVE button PRINT button and DEL button to enter corresponding working state directly 4 Set function User should check the system of the meter before using including system time types of temper...

Page 8: ...er to dia4 Diagram 4 Diagram 4 Press MONTH button repeatedly to select month you need press SETUP button to enter or use entering method press button spring out a entering window enter your needed mon...

Page 9: ...TUP button select item Default Temp press ENTER button to enter manual setting temperature Press SETUP button or direct press it once in the area of manual temperature setting spring out an input wind...

Page 10: ...electrode at 0 is not exactly 1000 At this time in order to ensure measuring precision it is neccessary to reset R0 value according to nominal resistance value Press SETUP R0 key and input R0 value o...

Page 11: ...standard buffer solution group 5 2 Setting standard buffer solution group The meter has automatic distinguishing function which can distinguish 10 different standard buffer solutions Multi point not...

Page 12: ...avoid overlapping of pH value user should select corresponding standard buffer solution to calibrate User should delete all the other selection of solution which are not used When selection finished p...

Page 13: ...e slope set as 100 potential value and temperature value In the middle of the screen it shows the current standard buffer solution group set by user At the bottom of the screen it shows the result of...

Page 14: ...iagram 12 if user need to calibrate some other standard buffer solution please press CONTINUE button then repeat above procedure till calibration finished Diagram 12 During the calibration user can fi...

Page 15: ...springs out a window and ask Save current result Press ENTER button to store Press CANCEL button to cancel The meter returns to starting working state 8 Print communication function Only one socket or...

Page 16: ...delete a certain stored data At VIEW Stored Datum mode press DEL button to select Delete current one press ENTER button to delete Press CANCEL button to exit the operation Note You should be very care...

Page 17: ...9 4 060 6 833 55 1 715 3 554 4 075 6 834 60 1 723 3 560 4 091 6 836 70 1 743 3 580 4 126 6 845 80 1 766 3 609 4 164 6 859 90 1 792 3 650 4 205 6 877 95 1 806 3 674 4 227 6 886 CLASS DENANARE 1th GRADE...

Page 18: ...E KHC4H4O6 25 SATURATION 7 3 776 1 GRADE KH2C6H5O7 0 05 11 41 4 008 1 GRADE KHC8H4O4 0 05 10 12 6 865 1 GRADE KH2PO4 0 025 3 387 Na2HPO4 0 025 3 533 7 00 DENANARE KH2PO4 0 02 2 780 Na2HPO4 0 0275 3 88...

Page 19: ...t 2 E 201D PH Combination Electrode 1 3 T 820D Temperature Sensor 1 4 Buffer Solution 4 7 10 5 bags for each 5 Multifunction Electrode Support Stand 1 6 Fuse 0 5A 5 7 AF30 Stylus 1 8 TP 16 Print Conne...

Reviews: