INE-PHS-3E 2016 

 

9

 
the solution even, pH value of solution will be displayed on screen. If temperature 
electrode is not connected to the meter, measuring procedure differs on whether 
temperature of measured solution is the same as the calibrated solution. The 
operating procedure is as follows: 
(1) 

When temperature of measured solution and temperature of positioning 

temperature are same, measuring procedure is as follows: 

a) Rinse the electrode head with distilled water, then clean it with measured 

solution; 

b) Insert electrode into measured solution, stir the solution with glass stick to make 

the solution even, pH of solution will be displayed on screen. 

(2) When temperature of measured solution is different from temperature of 

positioning solution, measuring procedure is as follows: 

a) Rinse the electrode head with distilled water, clean it with measured solution 

again; 

b) Measure temperature value of measured solution with thermometer; 
c) Press "

Ÿ

" or “

ź

” “Temp” button to make the meter display temperature value of 

measured solution, then press "ENTER" button; 

d) Insert electrode into measured solution, stir the solution with glass stick to make 

the solution even, pH value of the solution is displayed on screen. 

 

4.3    Electrode potential (mV) measurement 

1)  Turn on the meter to enter mV measurement state.   
2)  Clip ion selective electrode (or metal electrode) and reference electrode on the 

electrode support stand; 

3)  Rinse the electrode head with distilled water, clean it with measured solution once 

more; 

4)  Insert the plug of ion electrode into measuring electrode socket (6); 
5)  Insert reference electrode into reference electrode terminal (7) on back panel of 

the meter; 

6)  Insert above two electrodes into measured solution, stir the solution to be even, 

the electrode potential (mV value) of this ion selective electrode will be displayed 
on screen, ±polarity will also be displayed automatically; 

7)  If measured signal is over the measuring range of the meter, or measuring 

terminal is broken circuit, the screen won't be bright, it shows over load alarming 
Err. 

8)  When using metal electrode to measure electrode potential. Insert Q9 plug into 

measuring electrode socket (6), connect the clip and leading line of metal 
electrode; or using electrode transducer, one end of the electrode transducer is 
connected with electrode socket (6), metal electrode is connected with transducer. 
Reference electrode is connected with reference electrode terminal (7). 

 

 4.4 Temperature measurement 

The meter displays the temperature value directly in the any state of measurement 
when temperature sensor is connected with the meter. 
 

Summary of Contents for INE-PHS-3E

Page 1: ...INE PHS 3E pH Meter Operating Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 12 16...

Page 2: ...be used until it is up to standard z The warranty of the glass PH electrode is one year The electrode should be replaced in time after one year no matter whether it is used or not since the performan...

Page 3: ...INE PHS 3E 2016 2 CONTENT 1 GENERAL 2 MAIN SPECIFICATIONS AND PERFORMANCE 3 CONSTRUCTION 4 OPERATION INSTRUCTION 5 MAINTENANCE 6 CAUTIONS FOR ELECTRODE OPERATING 7 COMMON PROBLEMS AND SOLVING...

Page 4: ...MANCE 2 1 Range pH mV Temp Measuring range 0 14 00 pH displaying range 2 00 18 00pH 0 1999 mV automatic polar display 0 99 9 Resolution 0 01pH 1mV 0 1 Temp compensation 0 99 9 2 2 Accuracy pH mV Temp...

Page 5: ...alue larger manually at the pH measurement state Press the button to adjust the calibration slope pH value larger manually at the calibration slope adjusting state button This button is a reducing val...

Page 6: ...Enter slope adjusting state Press pH mV button If the temperature electrode is not connected to the meter enter manual adjusting temp state Press or button to adjust temp larger or smaller manually a...

Page 7: ...her calibration that it is not PH 4 00 PH 7 00 and PH 10 01 PH buffer Adjust PH value in the state of calibration to the PH value of the standard solution at this temperature to press Enter button and...

Page 8: ...ion press pH mV button to exit Two point calibration g Prepare two standard buffer solutions Insert the electrode which has been cleaned with distilled water into pH4 00 press Std button once the scre...

Page 9: ...will display Std yes Press Enter button to enter calibration state press any button to exit and return the measuring state Enter the calibrating state the meter recognizes the buffer automatically an...

Page 10: ...on screen 4 3 Electrode potential mV measurement 1 Turn on the meter to enter mV measurement state 2 Clip ion selective electrode or metal electrode and reference electrode on the electrode support s...

Page 11: ...lectrode glass bulb must not be touched to hard substance any damage will make the electrode cease to be effective 3 After measurement put on the electrode sleeve that has little supplementary liquid...

Page 12: ...n liquid interface is the main cause of measuring error 3 when measured solution is PH 10 00 please select buffer solution pH 10 01 4 when measured solution is PH 4 50 please select buffer solution pH...

Page 13: ...ature electrode E 201 PH sealed combination electrode Optional 0 14PH E 201 C PH refillable combination electrode Optional 0 14PH E 201 Z PH cone combination electrode Optional 0 14PH E 201 P PH plate...

Page 14: ...asuring state the instrument can not switch to pH measuring state When the measuring range is over 20 00pH the instrument will show Err first and then switch to mV measuring state 1 Reduce the input s...

Page 15: ...issolve 3 387g GR monopotassium phosphate and 3 533g GR disodium hydrogen phosphate in 1000ml high purity deionized water 3 pH 9 18 solution Dissolve 3 80g GR sodium tetraborate into 1000mL high purit...

Page 16: ...E pH Meter Number Description Quantity 1 INE PHS 3E pH Meter 1 set 2 Power Line universal 1 3 Buffer Solution pH 4 7 10 5 bags for each 4 Fuse BGXP 5 20 0 5A 2 5 Protective Cover small 1 6 PH Combinat...