2.1

 

Normal operating conditions: 

(1) Ambient temperature:                       5

~35

 

(2)Relative humidity:                             Not more than 80% 

(3)Temperature of measured solution:  5

~60

 

(4)Power supply:                                   two pieces of 5V batteries 

(5)No obvious vibration 

(6) No strong electromagnetic interference except the geomagnetic field. 

3. Construction

1-

 

Electrode socket    

2-

 

Electrode rack      

3-

 

Screen    

4-

 

Power ON/OFF 

5-

 

Increase key 

6-

 

Reduce  key 

       7 -  Mode key 

 8 -  Confirm key 

 9 -  Background light 

4. Operation

The  meter  works  on  two  pieces  of  5V  battery.  Open  the  battery  cover  on  the 

backboard of the meter and insert the battery.  

Please  calibrate  before  measuring  pH  value.  No  need  to  calibrate  in  each 

operation.  In  general,  calibration  is  needed  once  a  day  when  the  measuring 

interval is comparatively short.  

The meter has two points calibration. The procedure is as following: 

4.1

 

Press  “ON/OFF”  button  to  switch  on  the  battery  supply.  the  meter  enter 

“mV”  measuring state; 

4.2

 

Press  “MODE”  button  to  enter  temperature  measuring  state,  the  “

” 

indicating light flashes. The model circuit is as following: 

mV measurement (displaying MEAS and mV)        

temperature setting (the “

” light is flashing)  

pH  calibration 1(displaying STD1)    

pH calibration 2(displaying STD2)     

pH measuring (displaying MEAS and pH) 

4.3

 

Press  “

”  or  “

”  button  to  adjust  temperature  value  when  the  displayed 

temperature  value  is  the  same  as    the  temperature  value  of  calibrated 

solution,  press  “ENTER”  button  to  store  it  into  the  meter.  The  “

” 

indicating light stops flashing at this time. 

4.4

 

Press  “MODE”  button,  the  meter  display  “STD1”  to  enter  the  first 

calibration  point  (if  you  don’t  need  calibration,  press  “MODE”  two  times, 

the meter display “MEAS” enter PH measurement directly); 

Summary of Contents for INE-PHB-4

Page 1: ...INE PHB 4 Portable pH Meter Operation Manual...

Page 2: ...omatically It is suitable for measuring the pH value and potential mV in the laboratories and inspection offices The compact size and portable design of the meter make it specially suitable for being...

Page 3: ...comparatively short The meter has two points calibration The procedure is as following 4 1 Press ON OFF button to switch on the battery supply the meter enter mV measuring state 4 2 Press MODE button...

Page 4: ...as been used to measure strong acid pH 2 or strong alkali pH 12 The meter has been used to measure the liquid containing fluoride or solution with pH 7 or dense organic solutions When the temperature...

Page 5: ...minated clean it with medical cotton or with 0 1mol L diluted hydrochloric acid 5 4 If the end of the electrode is found to have cracks or become aged stored for over 2 years it is necessary to change...

Page 6: ...Press MODE button Press MODE button Press ENTER button Press MODE button Press ENTER button Press MODE button Press MODE button Press MODE button Packing list INE PHB 4 Portable pH meter NO d Descrip...

Page 7: ...trode Instruction Manual Characteristics One in two pH combination electrode INE E 201 Serials plastic shell polycarbonate INE 65 1 Serials Glass shell INE E 202 Serials Glass shell One in three pH co...

Page 8: ...After using the electrode please rinse it carefully 7 The components which touch sample solution include PC polycarbonate shell glass accessories and silicon rubber Before measurement you should confi...

Page 9: ...l solution 3 mol L for 4 hours 2 Adhered with organics please immerse the electrode in absolute ethyl alcohol or solution which can dissolve this organics for 15 minutes Stir solution with electrode b...

Reviews: