2.6.3
Check calibration parameters
The meter supports automatic identification function of the conductivity solution,
the default four JJG standard solution, standard data using potassium chloride
solution, nominal concentration respectively is 1 mol/L , 0.1 mol/L, 0.01 mol/L,
0.001 mol/L. If user adopts non-standard solution for the meter, set the
recognition type as "Manual identification", and manually enter the nominal value.
Check Measurement Parameter Diagram (Figure 10)
2.6.4 Start calibration
Specific calibration process is as follows:
Prepare 1~2 kinds of standard solution according to the need (Options);
Insert the conductive probe and ATC probe into corresponding measuring
electrode socket; clean the probes with DI water and put in the standard
solution;
If it is automatic recognition, the meter will automatically identify the solution,
and display the standard value; If the recognition is failure, check whether
the ATC probe is properly connected, whether the conductive probe is
damaged, or whether the shelf life of standard solution is over.
If it is manual identification, to manually input the nominal value.
If the solution cannot be identified or no input nominal value, this calibration
can not continue.
After reading is stable, press the "Calib" key, the meter store calibration
results and switch to report page, shows the calibration results;
If user need to continue to calibrate other standard solution, click "Continue
Calibration" button to return to calibration status. Clean the probes and put
in another standard buffer. After successful recognition by the meter, and
the reading is stable, press