Warning!
Forbid to transfer the vessel when the
instrument working.
5.
Trial run
•
Make sure the required operating voltage and power supply
voltage match.
•
Ensure the socket properly grounded.
•
Plug in the power cable
,
ensure the power is on and begin
initializing.
•
Add the medium into the vessel with an appropriate
stirring bar.
•
Place vessel on the work plate.
•
Set the target stirring speed and begin.
•
Set the target heating time.
•
Set the target temperature and start heating.
•
Stop the heating and stirring functions.
If these operations above are normal, the device is ready to
operate. If these operations are not normal, the device may be
damaged during transportation, please contact your local
dealer /supplier.
6.
Working with external temperature
sensor
Connection PT1000
Fuse
Figure 7
The external temperature sensor PT1000 is manufacture’s
standard accessory. If the sensor is plugged in, “Probe” will be
shown on the LCD digital display to indicate the sensor is
operating.The setting value of external temperature sensor and
actual temperature are displayed. Safe circuit controls hotplate
temperature.Comparing with the temperature control of the
hotplate, the external temperature sensor can control the
medium’s temperature more precise. The heating function will
be stopped automatically under abnormal conditions. Please
operate follow the instructions below:
•
Switch OFF the instrument.
•
Ensure the external temperature sensor is inserted in the
media heated.
•
Switch ON the instrument and run heating function.