background image

1

- Function knob

2, 2a  - Dual knob: 2 - Gas adjustment

2a - Air adjustment

3

- LED Standard

4

- LED StartStop

5

- LED Button

6

- LED Burner head HOT / BHC

7

- Burner head

7a - Flame orifice

8

- Connector for foot pedal (Cable)

9

- Power connector for 9 V DC

10 - Gas inlet R 1/4“ L (left hand thread)
11 - Holding device for inoculation loop holder

12 - Burner head screw
13 - Monitor electrode
14 - Ignition electrode

15 - Active nozzle
16 - Nozzle holder for alternative gas
17 - Cover of the burner shaft
18 - Retaining screw for

cover of the burner shaft

18a- Position scew for the cover
19 - Guide slots for tilt adjustment

4

6

2a

2

5

7

1

7a

3

14

13

12

8

9

10

11

16

17

18

15

19

19

18a

20 - Function LED and programming key

21

- Battery compartment (2 x AA, Mignon)

21a - Transport insulation for the batteries

21

20

21a

Summary of Contents for Fuego SCS basic

Page 1: ......

Page 2: ...vice for inoculation loop holder 12 Burner head screw 13 Monitor electrode 14 Ignition electrode 15 Active nozzle 16 Nozzle holder for alternative gas 17 Cover of the burner shaft 18 Retaining screw for cover of the burner shaft 18a Position scew for the cover 19 Guide slots for tilt adjustment 4 6 2a 2 5 7 1 7a 3 14 13 12 8 9 10 11 16 17 18 15 19 19 18a 20 Function LED and programming key 21 Batt...

Page 3: ...hose clamps are required All gas connections must be adequately tightened with two wrenches Ensure gas proofness with a suitable test fluid equipment DO NOT seal the swivel nut with Teflon tape etc Keep hands or other parts of the body away from the burner orifice 7a Do not operate the unit near flammable liquids or hazardous materials Unattended operation of the unit is not permissible Always wor...

Page 4: ... the same way and exchange the nozzles ATTENTION Before mounting a nozzle check the O Ring 20 Replace the sealing if damaged or worn Art No 8 000 010 Now you are ready to connect the gas supply to the gas inlet 10 The correct pressure for natural gas is within the range of 18 25 mbar for propane butane gas 28 57 mbar Only use DVGW or other gas approved safety tubings with thread or tubing connecto...

Page 5: ...switched off The radio system can be used without changing any further settings as the laboratory gas burner and the wireless foot pedal have already been programmed to each other gas burner and foot pedal programming see paragraph 2 4 by the manufacturers Attention The laboratory gas burner is started as soon as the foot pedal is operated and the LEDs 3 and 4 stop blinking Attention Flame The cor...

Page 6: ...gas burner 4 Shortly afterwards press the programming key 20 on the foot pedal for approx 1 second to start the programming procedure The function LED 20 will now start to glow This procedure can last for between 0 5 and 30 seconds As soon as the laboratory burner and the foot pedal have been successfully programmed the glowing is switched off and the function LED 20 switches to brief faint flashe...

Page 7: ... 18 25 mbar Nozzle P propane butane gas 47 57 mbar Flame failure Green LED Button Standard or StartStop blinks 3x This signal indicates a malfunction if the flame is extinguished by external factors and fails to reignite within 5 sec In case of flame failure check the burner head 7 for possible clogging and verify the correct input pressure of the gas supply In case of this malfunction the gas sup...

Page 8: ...ion is performed in the reverse sequence The dismounted burner head can be even dismantled into the individual components for in depth cleaning Unscrew both screws 12a and take off the base plate 12b of the burner head which was fixed by the two screws 12a After the base plate is removed both electrodes can be pulled out for seperate cleaning Reinstallation is performed in the reverse sequence 5 2...

Page 9: ... batteries Change both mignon batteries AA in the battery holder in the wireless foot pedal 21 Pay attention to correct insertion in accordance with the batteries polarity see label in the battery holder of the foot pedal 9 Warranty All Fuegos are covered under our two year manufacturer warranty against any manufacture defects in material and workmanship The 05 warranty guarantees all Fuegos under...

Page 10: ...Notes ...

Page 11: ... cannot encircle the electrodes correctly Clean this areas with a brush The burner head can be cleaned with customary commercial disinfectants or it can be sterilized in an autoclave or washed in a dishwasher In operating mode Button or Pedal StartStop flame burns 30 seconds only BHC time limit is active LED Burner head hot BHC is blinking The burner head is clogged and must be cleaned see paragra...

Page 12: ...his message indicates that the burner head is removed or not mounted correctly Reinstall burner head and reset the unit by a long push on the function knob 1 see paragraph 4 Amber LED Burner head HOT BHC is on permanently Residual heat display is active Attention DO NOT TOUCH the burner head see paragraph 3 LEDs Standard 3 and Start Stop 4 are blinking alternately Wireless foot pedal out of range ...

Page 13: ...es Flame temperature 1350 C on liquid gas 1300 C on natural gas E Temperature threshold level 1 kW liquid gas 1 kW natural gas Electrical Power consumption 2 VA Power connection 100 240 V 50 60 Hz max 0 3 A 9 V DC 1 A Power supply foot pedal 2 x AA Mignon Mechanical Casing and operating controls stainless steel glass UV and solvent resistant Burner head removable and decomposable stainless steel C...

Page 14: ...rd ElectricalEquipment classB table4 GenericImmunityStandard Industrialareas continuous unmonitored operation 1 2 EN61000 3 2 2000 Limitsforharmoniccurrentemissions classA 1 3 Protectionrequirementsconcerningelectromagneticcompatibility 3 1 2 Article3 1 b Electromagnetic compatibility and Radio spectrum Matters ERM ElectroMagnetic Compatibility EMC standard for radio equipment and services EN30148...