Brief, intense flashes at
rapid set intervals:
The RF stick is out of range or has been
switched off. If no burner is detected
after 40 flashes, the foot pedal
switches itself off.
Intermittent lighting up
of the LED, duration approx. 1 sec.:
batteries low
(see paragraph 8)
Slight glow:
Programming in progress
(see paragraph 2.4)
2.4 Foot pedal programming:
The laboratory burner and foot pedal can be newly programmed to each other if the signal
is interrupted or in cases of parallel use of several radio systems. For programming of the
radio channel, place the laboratory gas burner close to the wireless foot pedal (max. 30
centimetres). Programming has a restricted range to avoid inadvertent re-programming of
other burners. (In cases where undesired re-programming of a neighbouring burner occurs,
switch it off during the programming procedure.)
- Programming procedure:
Attention: Adhere to the sequence for the programming procedure. Close the gas
regulator (2) by turning it clockwise to avoid inadvertent activation of the flame.
1. Switch off laboratory gas burner.
2. Switch on foot pedal by operating briefly or press again briefly.
3. Switch the laboratory gas burner.
4. Shortly afterwards, press the programming key
(20)
on the foot pedal for approx. 1 second
to start the programming procedure. The function LED
(20)
will now start to glow. This procedure
can last for between 0.5 and 30 seconds. As soon as the laboratory burner and the foot
pedal have been successfully programmed, the glowing is switched off and the function
LED
(20)
switches to brief, faint flashes.
Note: The laboratory burner and the foot pedal have already been programmed to each
other by the manufacturers.
2.5 Operation: Application programs
- BUTTON StartStop:
The flame is ignited by operation of the function knob
(1)
. The flame
is extinguished after renewed actuation of the function knob
(1)
.
- PEDAL Standard:
The flame is ignited by operation of the foot pedal. The foot pedal remains depressed for the
duration of use. The flame is extinguished upon release of the pedal.
- PEDAL Start-Stop:
The flame is ignited by operation of the foot pedal. The flame is extinguished after renewed
actuation of the foot pedal. Additionally the flame is automatically extinguished when the
burning timer has expired after 60 min. Alternatively the flame can be extinguished by a short
push on the function knob.
2.6 Operation: Switch-off
The unit can be turned off by pushing the function knob
(1)
for more than 2 seconds.
The foot pedal is deactivated automatically.