10. Trouble Shooting
Problem
Possible cause
Solution
Piston overflows with
liquid
Piston wears out.
Contact with
manufacturer
Piston moves with
difficulty
Piston or its parts are
contaminated or
damaged due to
crystallization and
sedimentation.
① Do “Cleaning the
Barrel”
② Contact with
manufacturer
Failure to filling
Filling valve is clogged.
① Replace filling
valve
② Contact with
manufacturer
Failure to refill;refilling
sucks back into the
dispensing tip.
Dispensing valve is
contaminated or
dispensing tip
damaged.
Bubbles in the
instrument/Dispensing
volume is less than
the volume displayed.
Filling pipe is loose or
damaged.
Replace filling pipe
Filling pipe is away
from the liquid.
Checking filling pipe
Return pipe is not
installed or wrongly
installed.
Contact with
manufacturer
The instrument is not
fully refilled.
Checking
Operation
Filling valve is clogged
or damaged
Checking filling valve
Replace filling valve
No display
Battery dead
charging
Connection fail
USB cable
connection checking