background image

 

 

B.  Interpupillary adjustment of head. (Only for models B1-220 and B1-223) 

 

1.  Look through the eyepiece and adjust the distance between the eyepiece tubes (14) by 

grasping the head, by its four gripped corners (13), and moving it horizontally. 

 

2.  When a full field of view is observed through both lenses, and images blend into one, the 

interpupillary distance is correct for the user’s eyes. Take note of the index reading marked 
in the interpupillary scale (2) for the following step. 

 

3.  Adjust the diopter scales (14) situated on both eyepiece tubes until the reading of both 

scales coincides with the value of the interpupillary distance scale. This step is necessary to 
maintain the parfocality of the objectives.  

 

4.  Readjust the diopter scales for variation in interpupillary distance.  
 

C.  Focusing the microscope. 

 

1.  Turn the revolving nosepiece (3) and place the 4X objective (4) in the optical path ensuring 

that it is properly clicked into place.  

 

2.  Turn the coarse focus knob (19) until the stage (8) is in its lowest position.  

 

3.  Place a microscopic sample on the stage, make sure the cover slip faces upwards.  

 

• 

Swing the moveable finger (5) on the mechanism (6), place specimen against fixed side 
of slide holder (7) and gently release the moveable finger until the specimen is 
supported in place. 

 

4.  Make sure that the specimen is in the optical path by adjusting the stage using the knobs 

controlling the X/Y movements (21). 

 

5.  Looking through the eyepiece, turn the coarse focus knob until the preparation appears in 

focus.   

 

6.  Readjust the focus with the fine focus knob (20) until the image appears sharply defined.  

 

D.  Adjust diopter for differences in eyesight. (Only for models B1-220 and B1-223) 

 

1.  With your right eye, look into the right eyepiece tube (1) and adjust the sharpness of the 

image using the fine focus knob (20). 

 

2.  With your left eye, looking through the left eyepiece, adjust the focus by rotating the diopter 

corrector (14) on the left hand eyepiece tube until the image is sharp. Do not use the fine 
focus knob.  

 

E.  Adjusting the aperture of the diaphragm. 

 

The aperture of the diaphragm (9) should not be used to control the brightness of the light. 
Diaphragms are designed to obtain high resolution of the sample, and provide contrast in the 
image. Smaller apertures will deliver higher contrast to image. However, closing the aperture 
too much will reduce resolution. Experimentation is the best method of determining the 
correct aperture of the diaphragm. Suggested apertures for each objective are as follows:  
 
  

OBJECTIVE 

 

APERTURE OF IRIS 

 
  

4X 

 

 

From fully closed to 1/8 open. 

  

10X 

 

 

From 1/8 to 1/4 

  

40X 

 

 

From 1/4 to 1/2 

  

100X 

 

 

From 1/2 to 3/4 

 

 

-6- 

Summary of Contents for B1-220ASC

Page 1: ...PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 02 11 10 INSTRUCTION MANUAL FOR B1 220ASC Biological Microscope...

Page 2: ...echanism 7 Slide holder 8 Stage 10 Condenser 9 Diaphragm 15 Head 17 Arm 14 Eyepiece tube with diopter correctors 18 Knurled locking screws for securing specimen holder 19 Coarse Focus Knob 20 Fine Foc...

Page 3: ...the series Unpacking All components of the microscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case yo...

Page 4: ...the focussing knob on the condenser Fig 1 until the condenser holder reaches its lowest position Unscrew the knurled locking screw on the condenser and insert the condenser in the mounting ring so tha...

Page 5: ...microscope Firmly re tighten the knurled locking screw E Eyepieces 1 Remove the protective covers on the eyepiece tubes 14 Gently place the eyepieces in the tubes without touching the lens surface B1...

Page 6: ...lease the moveable finger until the specimen is supported in place 4 Make sure that the specimen is in the optical path by adjusting the stage using the knobs controlling the X Y movements 21 5 Lookin...

Page 7: ...must ONLY be used with the 100X objective which is the only objective prepared for it If any other objective comes into contact with immersion oil it must be cleaned immediately G Critical illuminatio...

Page 8: ...otective cap that covers the port can be removed Fig 4 Adapter tube Light controlling rod Knurled screw Fig 3 Photo adapter Video adapter 3 Insert the adapter tube with the camera mounted on the verti...

Page 9: ...OCEDURE NOT COVERED IN THIS INSTRUCTION MANUAL A Optical maintenance 1 Do not attempt to disassemble any lens component 2 Before cleaning any lens remove surface dust with a fine brush especially for...

Page 10: ...s Fig 5 g Close hinged door and tighten locking screw 2 Changing the fuse a Using a flat screwdriver gently press on the slot located on the fuse cap Fig 1 and make a turn in the direction indicated b...

Page 11: ...t of the lens should it come into contact with the slide An additional measure of security consists of an adjustable stopper on the ascending movement of the stage The stopper comes pre adjusted by th...

Page 12: ...by a qualified technician IMAGE QUALITY PROBLEM CAUSE SOLUTI N No image Nosepiece not positioned properly Image too bright Turn until clicks into place Reduce the intensity of the light Poor resoluti...

Page 13: ...sion Warrantee All MRC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of th...

Reviews: