a.  To tighten the focussing controls of the coarse focus knob, the ring must be turned in an 

anti-clockwise direction, as indicated by the arrow. To loosen it, the ring must be turned 
clockwise.  

 

Stage stopper 
screw 

Coarse focus 
tension control 

Condenser 
focussing 
control 

Fig. 6 

2.  Adjusting the stage stopper. 

 

The 40X and 100X (optional) objectives incorporate security measure in that the lens tip  
retracts to avoid damage to the front of the lens should it come into contact with the slide.  
An additional measure of security consists of an adjustable stopper on the ascending 
movement of the stage. The stopper comes pre-adjusted by the manufacturer for standard 
slides with cover slips 0.17mm thick. For observing other types of samples, however, it may 
become necessary to readjust the stage stopper.  

 

a.  Loosen the stage stopper screw (Fig. 6) with a 2mm hexagonal key. 

 

b.  Focus on the sample using only the fine focus knob (20) with firstly the 4X objective (4), 

and 10X objective. 

 

c.  Tighten the stage stopper screw until tight enough for the stage (8) not to move 

upwards.  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

-11- 

Summary of Contents for B1-220ASC

Page 1: ...PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 02 11 10 INSTRUCTION MANUAL FOR B1 220ASC Biological Microscope...

Page 2: ...echanism 7 Slide holder 8 Stage 10 Condenser 9 Diaphragm 15 Head 17 Arm 14 Eyepiece tube with diopter correctors 18 Knurled locking screws for securing specimen holder 19 Coarse Focus Knob 20 Fine Foc...

Page 3: ...the series Unpacking All components of the microscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case yo...

Page 4: ...the focussing knob on the condenser Fig 1 until the condenser holder reaches its lowest position Unscrew the knurled locking screw on the condenser and insert the condenser in the mounting ring so tha...

Page 5: ...microscope Firmly re tighten the knurled locking screw E Eyepieces 1 Remove the protective covers on the eyepiece tubes 14 Gently place the eyepieces in the tubes without touching the lens surface B1...

Page 6: ...lease the moveable finger until the specimen is supported in place 4 Make sure that the specimen is in the optical path by adjusting the stage using the knobs controlling the X Y movements 21 5 Lookin...

Page 7: ...must ONLY be used with the 100X objective which is the only objective prepared for it If any other objective comes into contact with immersion oil it must be cleaned immediately G Critical illuminatio...

Page 8: ...otective cap that covers the port can be removed Fig 4 Adapter tube Light controlling rod Knurled screw Fig 3 Photo adapter Video adapter 3 Insert the adapter tube with the camera mounted on the verti...

Page 9: ...OCEDURE NOT COVERED IN THIS INSTRUCTION MANUAL A Optical maintenance 1 Do not attempt to disassemble any lens component 2 Before cleaning any lens remove surface dust with a fine brush especially for...

Page 10: ...s Fig 5 g Close hinged door and tighten locking screw 2 Changing the fuse a Using a flat screwdriver gently press on the slot located on the fuse cap Fig 1 and make a turn in the direction indicated b...

Page 11: ...t of the lens should it come into contact with the slide An additional measure of security consists of an adjustable stopper on the ascending movement of the stage The stopper comes pre adjusted by th...

Page 12: ...by a qualified technician IMAGE QUALITY PROBLEM CAUSE SOLUTI N No image Nosepiece not positioned properly Image too bright Turn until clicks into place Reduce the intensity of the light Poor resoluti...

Page 13: ...sion Warrantee All MRC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through misuse or modification of th...

Reviews: