5
SECTION 3
INSTALLATION
Local city, county, or other ordinances may govern the use of this equipment. If you have
any questions about local requirements, please contact the appropriate local agency.
Installation may be performed by the end user.
Under normal circumstances these units are intended for use indoors, at room
temperatures between 5
°
and 40
°
C, at no greater than 80% relative Humidity (at 25
°
C)
and with a supply voltage that does not vary by more than 10%. These ovens should not
be operated at an altitude exceeding 2000 meters. Installation category is CAT-II Pollution
Degree 2. Customer service should be contacted for operating conditions outside of these
limits.
3.1 Power
Source:
The unit power requirements are listed on the data plate. PLUG
THE UNIT INTO A PROPERLY GROUNDED AND RATED RECEPTACLE OF
THE CORRECT STYLE.
THE VOLTAGE OF THE RECEPTACLE SHOULD NOT
VARY MORE THAN 10% FROM THE DATA PLATE RATING.
A separate circuit
is recommended for this unit to prevent loss of product due to overloading or circuit
failures caused by other equipment.
3.2 Location:
When selecting a site for the unit, consider conditions which may affect
performance, such as heat from radiators, ovens, autoclaves, etc. Avoid direct sun,
fast-moving air currents, heating/cooling ducts, and high-traffic areas. To ensure air
circulation, allow a minimum of 15cm between the unit and any walls or partitions
which might obstruct free air flow.
3.3 Lifting
/
Handling:
These units are heavy and care should be taken to use
appropriate lifting devices that are sufficiently rated for these loads. Units should
only be lifted from their bottom surfaces. Doors, handles and knobs are not
adequate for lifting or stabilization. The unit should be completely restrained from
tipping during lifting or transport. All moving parts such as shelves and trays should
be removed and doors need to be positively locked in the closed position during
transfer to prevent shifting and damage.
3.4 Leveling:
The unit must sit level and solidly. Model 1425 has four (4) rubber feet
that are already attached to the unit and are not adjustable. Leveling feet are
supplied with models 1445 and 1465 and must be installed in the four holes at the
base of the unit. With the unit standing upright, turn the leveling feet
counterclockwise to raise the level of that corner. Adjust each foot until the unit
stands solid and level. If the unit must be moved, turn the leveling feet in all the way
to prevent damage while moving.