22
Comfort Beep
With the Comfort Beep enabled, the buzzer beeps once each minute to indicate that all is OK and
the instrument is running normally.
Press the Up or Mode key to step from one option to the next. Then press Right to make a selection
followed by Left to save and exit.
Man-Down Alarm
The Man-Down alarm notifies nearby personnel if the instrument remains motionless for about 30
seconds when it should be moving, indicating a possible worker in distress. If the wireless option is
enabled, alarm notification is also sent to a remote location. During the Warning Time the alarm
beeps 2x per second and asks the user ‘Are You OK?’. If so, the user can acknowledge and stop
the alarm by pressing the Left Key, or simply tilting the instrument briefly. If the Warning Time
expires without acknowledgment, the full Man-Down Alarm of 3x beeps per second begins.
Man-Down Alarm Off/On
Press Right to make a selection followed by Left to save and exit.
Sensitivity
This sensitivity of motion sensor can be set to Low, Medium or High. Low
sensitivity means fewer alarms and High sensitivity means easier detection of
stopped motion and more chance of an alarm. Use Up or Mode to move up or
down the list, Right to select, and Left to save and exit.
Warning Time
The Warning Time is the amount of time the Man-Down alarm stays on after it
is triggered. The default time is 30s.
Datalog
The instrument automatically stores the concentration readings at regular time
intervals (this function cannot be turned off). In the datalog sub-menu, a user
can adjust the interval or clear all data. Press Right to enter the Datalog menu.
Clear Datalog
This erases all the data stored in the datalog.
Caution:
Once the data are cleared, they cannot be recovered.
O
L o w
M e d i u m
O
H i g h
Sensitivity
Datalog
Clear Datalog
Interval
Datalog
Clear Datalog
Interval
Set Montionless
_ _ 3 0 s
√ ×
Set Warning Time
O
O f f
O n
Off/On