After being given a
DATA:ISTA:VERB
command, the apparatus transmits the samples
in the following format:
Fstart
<start_value>
Hz,Fstop
<stop_value>
Hz,Fstep
<step_value>
Hz,N level
<levels number>
,dBuV/m=><0D><0A>
<data1>;<data2>;...<data n>
Where:
<start_value>
= string containing the Hz value of the start frquency
<stop_value>
= string containing the Hz value of the stop frequency
<step_value>
= string containing the step Hz value
<levels_number>
= string containing the number of samples transmitted below, that
represent the number of acquired levels
< N date > string containing the level at F frequency
In order to determine the F frequency =
<start_value>+N x <step_value>
Example for receive:
Fstart 9.750e+07 Hz,Fstop 1.025e+08 Hz,Fstep 5.000e+04 Hz,N level 101.000000,V/m
=>
0.02;0.03;0.02;0.03;0.02;0.02;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;
0.01;0.01;0.01;0.01;0.02;0.02;0.02;0.03;0.03;0.03;0.02;0.01;0.01;0.01;0.01;0.01;
0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;
0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.02;0.01;0.01;0.01;0.01;0.01;0.01;
0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;0.01;
0.01;0.01;0.01;0.01;0.01;0.01;0.02;0.01;0.01;0.01;0.02;0.02;0.02;0.01;0.01;0.01;
0.01;0.01;0.01;0.01;0.01;
Send this sequence:
1.
Command IDN:
Send string
*IDN?\r\n
2.
Configure the operating limits to define the capture window
a.
Stop frequency 102.6 MHz
send string :
SENS:FREQ:STOP:102.5e6\r\n
b.
Start frequency 97.6 MHz
send string:
SENS:FREQ:STAR:97.5e6
3.
Resolution bandwidth configuration (RBW)
Send string
SENS:BAND:300 KHz\r\n
4.
Detector type configuration