USER MANUAL
S E M S
- 36 -
PROPRIETARY REQUEST COMMANDS
Commands Description
CFG?
This command *
CFG?
Gives the serial number and the calibration date,
followed by [0D][0A]
Reception Example:
001_160616
where:
001 = serial number
_ = separation special character
160616 =
ddmmyy
calibration date
STA?
This command *
STA?
Gives the status of the machine in a string containing:
Vbat 4054mV;TBD;TBD;TBD;Ric 0;USB 1;Temp 0gr;Fun 0<0D><0A>
Where:
Vbat
xxxx
mV; => xxxx represent the voltage of the apparatus’ internal batteries
TBD => for future expansions
Ric
n
=> n if = 0 is not recharging, or:
if =1 the internal batteries are recharging
USB
n
=> n if = 0 USB cable not connected; if =1 USB cable connected
Temp
x
gr => x contains the °C apparatus’ internal temperature
Fun
c
=>c contains the fan status according to the following bit coding:
// ---1 if(
c
&& 0x01) =>Fun1, active; otherwise not
active
// --1- if(
c
&& 0x02) => Fun1, rotating; otherwise still
// -1-- if(
c
&& 0x04) => Fun2, active; otherwise not active
// 1--- if(
c
&& 0x08) => Fun2, rotating; otherwise still
PROPRIETARY IMPLEMENTATION COMMANDS
The commands towards the instrument are divided into multiple levels and separated
from
“:”
Semantic:
<level1>:<level2>:<level3>:<level4>\r\n
Summary of Contents for SEP
Page 16: ...USER MANUAL S E M S 16 c Link d Report e End Figure 6 Installer MPB SEP ...
Page 25: ...USER MANUAL S E M S 25 Figure 19 Manage Configurations ...
Page 44: ...USER MANUAL S E M S 44 Select Add a device Bluetooth Figure 33 Select RNBT xxxx ...