A n g l e Te s t O p e r a t i o n
Angle Set-Up
A. Angle Threshold
B. Angle Polarity
C. Angle Encoder
All angle tests must be entered using the PC Windows based Calibration Program. Once this is done the tests can be
accessed using the soft key on the PTT.
Entering data for the Angle Test is done using the PC Windows based Calibration Program. See the Screen above.
Select the “Angle Tests” tab. Enter the relevant information on this screen as needed for type of angle test that will be
used.
Note:
* The Bolt Audit Mode is an additional feature that is only available with PTT Bolt Audit model (Item # 072995)
Angle Mode
A. Angle Threshold
B. Torque Threshold
C. Allow Angle Rachet
Bolt Audit Mode
A. Initial Torque Threshold
B. Final Torque Threshold
Spindle Audit Mode
A. Initial Torque Threshold
B. Low Angle Threshold
C. High Angle Threshold
Bolt Audit Explanation
This feature is designed for auditing the torque applied to a fastener for a test such as determining relaxation after a fas-
tener has been tightened. An initial and final torque threshold is entered into the bolt audit mode fields. As an example
the initial torque value could be 70% of the desired torque on the fastener and the final torque would be the nominal
value for which the fastener was previously tightened. In this case, if the nominal torque is 500 lbf.ft, the initial torque
threshold would be entered as 350 lbf.ft, 70% of the nominal value.