EZ-TorQ-III User Manual
mountztorque.com - 1080 N 11
th
St - San Jose CA 95112 - 408.292.2214
10
6.
Set Up
6.1.
Setting
No
Item
Description
1
Up/Down &Numeric key pad
Target torque can be selected by the UP and DOWN arrow. Touch the center of
target torque to open the numeric key pad.
2
Target Torque & Tolerance
Check on/off to apply the target torque and tolerance settings.
3
Auto clear (sec)
Select amount of time (seconds) to clear the test data displayed. Or disable to
manually clear data
4
Torque Unit
Torque unit of measure (Kgf.cm, Kgf.m, Nm, cNm, Ozf.in, Lbf.in, Lbf.ft)
5
Mode
Peak torque, Track torque or First-peak torque
6
LPF (Low Pass Filter) freq
LPF (Low Pass Filter) frequency setting. ISO standard is 500Hz
7
Direction
Rotation direction selection Clockwise, Counterclockwise or Both directions
8
Tool SN
You can enter the serial number of the tool you want to measure.
This serial number is stored on the SD card along with the test