MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE
FIGURE 17:
Connect the main connector (A) to the main connector socket in the Workstation (B). Place the irrigation
tubing (C) into the irrigation pump marked with blue markers (D) between the marked blue lines of the tubing, then
close the irrigation pump head. Remove the irrigation tube spike’s cap (E), then either connect the irrigation tubing
to the irrigation bag (F) or place the spike in a bottle (G) and screw the provided water bottle cap (H) onto the
bottle securely. Ensure that the irrigation line clamp (I) is open. Ensure all tubing is free of any kinks.
4.2.2. Attaching the Umbilical Section to the Workstation
The Workstation should be cleaned prior to attaching the Umbilical Section to the Workstation by
thoroughly wiping down the Workstation with a viricidal cloth (Opti-cide, Cavicide or similar surface
disinfectant of choice) that has an active ingredient of isopropyl alcohol (17% - 21%).
Remove the Umbilical Section from the tray. To attach the Umbilical Section to the Workstation complete
the following steps:
FIGURE 18:
Connect the suction line to the waste container (A) and ensure that at least one port, in addition to the
vacuum port (B), is open on the container or attach the suction line to a waste bag (C). Place the suction tubing (D)
into the suction pump (E) marked with a single black and double blue lines aligning the black line of tubing with
black line on pump head and blue lines on pump head to blue lines on tubing. Then close the suction pump head.